T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2022-52509
- Event Type
- Malfunction
- Date Received
- June 1, 2022
- Date of Event
- May 2, 2022
- Report Date
- May 9, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00853052007257
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
PER TANDEM USER GUIDE: CHANGE YOUR INFUSION SET EVERY 48¿72 HOURS AS RECOMMENDED BY YOUR HEALTHCARE PROVIDER CHANGE YOUR CARTRIDGE EVERY 48¿72 HOURS AS RECOMMENDED BY YOUR HEALTHCARE PROVIDER. ALWAYS REMOVE ALL AIR BUBBLES FROM THE CARTRIDGE BEFORE BEGINNING INSULIN DELIVERY. ENSURE THERE ARE NO AIR BUBBLES WHEN DRAWING INSULIN INTO THE FILLING SYRINGE, HOLD THE PUMP WITH THE WHITE FILL PORT POINTED UP WHEN FILLING THE TUBING, AND ENSURE THAT THERE ARE NO AIR BUBBLES IN THE TUBING WHEN FILLING. AIR IN THE SYSTEM TAKES SPACE WHERE INSULIN SHOULD BE AND CAN AFFECT INSULIN DELIVERY NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. DEVICE NOT RETURNED.
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. REPORTEDLY, THE CUSTOMER WAS USING SUPPLIES BEYOND LABELING AND WAS NOT PERFORMING THE AIR REMOVAL PROCESS CORRECTLY. TANDEM TECHNICAL SUPPORT PROVIDED PROPER SUPPLIES LABELING AS WELL AS EDUCATED THE CUSTOMER ON THE PROPER AIR REMOVAL PROCESS. THE CUSTOMER REPLACED THE PUMP SUPPLIES AND RESUMED INSULIN THERAPY. CUSTOMER'S BLOOD GLUCOSE WAS IN THE RANGE OF 132-352 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1599464 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1000096 | 00853052007257 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Female | INFUSION SET: TRUSTEELINSULIN: NOVOLOG |