MONOPLUS C VIOLET 1 (4) 25M
Report
- Report Number
- 3003639970-2022-00224
- Event Type
- Malfunction
- Date Received
- June 1, 2022
- Report Date
- June 1, 2022
- Manufacturer
- B. BRAUN SURGICAL, S.A.
- Product Code
- NEW
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICE OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. UNDER PMA/510(K) NUMBER: K031216. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED ONE OPEN CASSETTE FOR ANALYSIS. THE DATE OF CASSETTE'S OPENING IS NOT WRITTEN. THREAD ON SAMPLE RECEIVED IS BROKEN INSIDE THE CASSETTE AND IT IS NOT USEFUL. THREAD WAS PROBABLY TANGLED IN THE REEL AND WHEN PULLED OUT FROM CASSETTE THE THREAD BROKE BECAUSE THE EXTRACTION WAS TOO HARD. REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFILL USP/EP AND B. BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: TAKING INTO ACCOUNT THAT THE CASSETTE RECEIVED DOES NOT FULFIL B. BRAUN SURGICAL SPECIFICATIONS, WE CONCLUDE THAT THE COMPLAINT IS CONFIRMED BY EVIDENCE OF THE FAILURE IN THE CASSETTE RECEIVED. ACTIONS ON DISTRIBUTED PRODUCT OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.
IT WAS REPORTED AN ISSUE WITH MONOPLUS C SUTURE. THE CLIENT (VETERINARIAN) REPORTED THAT, WHEN TRYING TO PULL THE THREAD OUT OF THE CASSETTE, THE THREAD BROKE INSIDE AND THEREFORE THE CUSTOMER WAS NOT ABLE TO USE THE PRODUCT. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1933636 | MONOPLUS C VIOLET 1 (4) 25M | SYNTHETIC ABSORBABLE MONOFILAR SUTURE | NEW | B. BRAUN SURGICAL, S.A. | F0024274 | 121035 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |