FDA Adverse Event Other Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1455912 · Received August 21, 2009

Report

Report Number
3004464228-2009-00722
Event Type
Other
Date Received
August 21, 2009
Date of Event
July 29, 2009
Report Date
July 29, 2009
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WILL NOT BE RETURNED SO NO EVAL IS POSSIBLE. THE CUSTOMER EXPERIENCED DIFFICULTY WHEN TRYING TO INJECT THE POD WITH INSULIN DURING THE FILL PROCESS WHICH INDICATES A PROBABLE PROBLEM WITH THE RETAINER THAT ALLOWED A COMPONENT TO MOVE RESULTING IN DAMAGE. THIS DAMAGE PREVENTED THE PLUNGER FROM ADVANCING. THE USER WOULD HAVE BEEN AWARE OF A PROBLEM DURING THE FILL PROCESS. THERE IS RESISTANCE FELT IN FILLING THE POD WITH INSULIN AND A DISTINCT "CRACKLING" NOISE RESULTING FROM STRIPPING THE THREADS OF THE COMPONENT WHEN OVER-PRESSURIZED (THOUGH NO "CRACKLING" NOISE WAS NOTED IN THIS REPORT). THE OMNIPOD USER GUIDE STATES: "WARNING: NEVER USE A POD IF, DURING FILL, YOU DETECT ANY CRACKLING NOISE OR RESISTANCE WHILE DEPRESSING THE PLUNGER OF THE FILL SYRINGE. USING A POD WITH THESE CONDITIONS COULD RESULT IN UNDER-DELIVERY OF INSULIN". THE USER IS ALSO INSTRUCTED IN THE USER GUIDE TO MONITOR BLOOD GLUCOSE LEVELS FREQUENTLY. BY FOLLOWING THESE RECOMMENDATIONS, THE USER BECAME AWARE OF THEIR HIGH BLOOD GLUCOSE AND STARTED A NEW POD OR BACKUP THERAPY IF NEEDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT HE HAD DIFFICULTY FILLING THE POD WITH INSULIN. DESPITE THIS, INSULIN WAS FORCED IN AND THE POD WAS PLACED. OVER THE NEXT TWO DAYS, HE EXPERIENCED SPORADIC HIGH BG'S (260-437 MG/DL). THE POD WAS DEACTIVATED BUT WILL NOT BE RETURNED FOR EVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L30140

Patients

Seq Age Sex Outcome Treatment
1 36 YR Other