FDA Adverse Event Malfunction Summary report: N

028066

MDR report key: 14552277 · Received May 31, 2022

Report

Report Number
9612330-2022-00033
Event Type
Malfunction
Date Received
May 31, 2022
Date of Event
March 28, 2022
Report Date
August 16, 2022
Manufacturer
NATUS MEDICAL INCORPORATED
Product Code
GWQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORT REF NATUS COMPLAINT# (B)(4) FOR 3RD SERIAL (B)(4). 3- BREAKOUT BOXES WITH CONNECTOR DAMAGE FROM ELECTRODE CONNECTOR DAMAGE: (1) SN: (B)(4) - BILLABLE, WARRANTY ENDED 3/14/2022 - SIEBEL DOESN'T SHOW ANY RECORD OF TROUBLESHOOTING BEING DONE TO CONFIRM FAILURE. BREAKOUT BOX WITH SN:(B)(4) EXCHANGE WAS DOCUMENTED UNDER SR (B)(4). (2) SN: (B)(4) - WARRANTY ENDS 9/27/2022. (3) SN: (B)(4) - WARRANTY ENDS 9/27/2022. TH ELECTRODE CONNECTORS GET MELTED INSIDE BREAKOUT BOX CHANNEL INPUTS. ITS DESCRIBED AS PLASTIC GETS MELTED INSIDE. WHEN THEY PRY TO REMOVE THE ELECTRODE CONNECTOR, THIS CAUSES PHYSICAL DAMAGE TO CONNECTOR CHANNEL OPENING(S). THE METAL PIN INSIDE (COPPER) IS DARKER (BLACKEN) FROM THE HEAT FORMING THE BLACK COPPER OXIDE. THIS HAPPENS EVERY 2 MONTHS. THEY HAD REPLACED SOME OF THEIR OLDER HARDWARE. THE CUSTOMER SAID SAME ISSUE HAPPENED ON SOME OLDER HARDWARE TOO. PICTURES WERE SENT - IT IS HARD TO SEE THE INSIDE OF THE CONNECTOR CHANNEL TO GET TRUE VIEW OF DAMAGE. FROM PICTURES YOU CAN DISTINGUISH BAD CHANNEL OPENING TO NON-DAMAGE. LOT/SERIAL NUMBER:(B)(4). INSTALL DATE: SEPTEMBER 28TH, 2021 AND SEPTEMBER 28TH, 2021. ACCEPTABLE RISK ASSOCIATED WITH THE COMPLAINT AS PER HAZARD ID LINE (B)(4) XLTEK EEG PSG RISK ANALYSIS SPREADSHEET. NO DEATH OR SERIOUS INJURY, CONSIDERED A CUSTOMER INCONVENIENCE. CUSTOMER'S ALLEGEDLY DEFECTIVE BREAKOUT BOX WAS RECEIVED. CURRENTLY WAITING FOR CUSTOMER'S RESPONSE TO ADVERSE EVENT QUESTIONNAIRE AND INVESTIGATION RESULTS FROM ENGINEERING.

Additional Manufacturer Narrative · 0

FOLLOW UP REPORT 001 REF NATUS COMPLAINT# (B)(4) FOR 3RD SERIAL (B)(6). PER QMS-004442 COMPLAINT HISTORIES ARE REVIEWED ROUTINELY PER QUALITY SYSTEM REQUIREMENTS AND ANY COMPLAINT TRENDS ARE ASSESSED AND DOCUMENTED AS PART OF THESE REVIEWS. DEVICE HISTORY WAS REVIEWED AND NO RELATED FAULTS/ISSUES FOUND. THE FOLLOWING INVESTIGATION RESULTS WERE DOCUMENTED: ENGINEERING INVESTIGATED THE ALLEGEDLY DAMAGED HEADBOX. INVESTIGATION DETERMINED "THE CUSTOMER THOUGHT THEY WERE SEEING BLACK COPPER OXIDE BEING FORMED. WHEN THE UNIT WAS DISASSEMBLED THERE IS NO RESIDUE OF ANY TYPE OBSERVED, AND THE GOLD PLATING WOULD NEED TO BE REMOVED FROM THE BRASS PIN FOR ANY TYPE OF CORROSION TO OCCUR. THE DARKER APPEARANCE OF THE ELECTRODE WITH THE TIP BROKEN OFF IN IT IS DUE TO THE LACK OF A REFLECTION FROM THE BASE OF THE PIN AT THE BOTTOM OF THE HOLE." THE ISSUE WAS DETERMINED TO BE CAUSED BY "IS AN ELECTRODE CONNECTOR TIP BREAKING OFF IN THE AMPLIFIER DUE TO BEING CRACKED/DAMAGED NEAR THE TIP AND BEING HELD IN PLACE BY A SUBSTANCE SUCH AS CONDUCTIVE GEL OR A CLEANING AGENT THAT WAS PRESENT IN THE AMPLIFIER CONNECTOR HOLE." CLOSURE RATIONALE: COMPLAINT VERIFIED, BEING TRACKED AS A TREND. LOW SAFETY RISK OF HARM, INDIVIDUAL COMPLAINT RELATED TO ISSUE STATED. COMPLAINT WILL BE INCLUDED IN TRENDING DATA FOR FURTHER REVIEW AND INVESTIGATION IF NEEDED.

Description of Event or Problem · 0

EMU40EXQ BREAKOUT BOX - HEADBOX CONNECTOR OPENING(S) HAS DAMAGE DUE TO BROKEN ELECTRODES INSIDE BOX. THE ELECTRODE CONNECTORS GET MELTED INSIDE BREAKOUT BOX CHANNEL INPUTS. NO INJURIES REPORTED.

Description of Event or Problem · 0

EMU40EXQ BREAKOUT BOX - HEADBOX CONNECTOR OPENING(S) HAS DAMAGE DUE TO BROKEN ELECTRODES INSIDE BOX. THE ELECTRODE CONNECTORS GET MELTED INSIDE BREAKOUT BOX CHANNEL INPUTS. NO INJURIES REPORTED.