L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM
Report
- Report Number
- 0001625425-2022-01014
- Event Type
- Malfunction
- Date Received
- May 31, 2022
- Date of Event
- May 17, 2022
- Report Date
- July 7, 2022
- Manufacturer
- ARGON MEDICAL DEVICES
- Product Code
- LJS
- UDI-DI
- 00886333209934
- PMA / PMN Number
- K091670
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
THE SAMPLE IS INDICATED AS AVAILABLE FOR RETURN. AS OF THE DATE OF THIS REPORT, THE SAMPLE HAS NOT BEEN RETURNED. A FOLLOW-UP REPORT WILL BE PROVIDED ONCE THE DEVICE HAS BEEN RECEIVED AND REVIEWED.
A REVIEW OF THE DHR AND INSPECTION RECORDS WAS CONDUCTED, AND NO SIMILAR CONCERNS WERE FOUND. ONE OPENED CATHETER WAS RETURNED FOR REVIEW. VISUAL INSPECTION CONFIRMED A CRACK IN THE LUER THAT WOULD RESULT IN LEAKAGE. SEVERAL COMPLAINTS FOR THIS PART NUMBER HAVE PREVIOUSLY BEEN RECEIVED REGARDING A CRACKED HUB RESULTING IN LEAKAGE, AND CAPA 2021-039 WAS INITIATED TO ADDRESS THIS ISSUE. THE CAPA IS CURRENTLY IN THE IMPLEMENTATION PHASE AND WILL EVALUATE THE CORRECTIVE ACTION IMPLEMENTATION FOR EFFECTIVENESS TO PREVENT A RECURRENCE OF THIS ISSUE. CAPA 2021-039 WAS INITIATED TO ADDRESS THIS ISSUE AND IS CURRENTLY IN THE IMPLEMENTATION PHASE. THE CAPA WILL EVALUATE THE CORRECTIVE ACTION IMPLEMENTATION FOR EFFECTIVENESS TO PREVENT A RECURRENCE OF THIS ISSUE.
PLASTIC CONNECTOR BETWEEN LINE AND EXTENSION SET CRACKED AND LEAKING. LOT#11401159 DATE OF DISCOVERY: MAY 17, 2022 THIS OCCURRED WHILE IN USE ON A PATIENT AND REQUIRED REMOVAL FROM THE PATIENT.
PLASTIC CONNECTOR BETWEEN LINE AND EXTENSION SET CRACKED AND LEAKING. LOT#11401159 DATE OF DISCOVERY: (B)(6) 2022 THIS OCCURRED WHILE IN USE ON A PATIENT AND REQUIRED REMOVAL FROM THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2442264 | L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM | L-CATH PICC | LJS | ARGON MEDICAL DEVICES | 384539 | 11401159 | 00886333209934 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention| O |