POWERFLEX P3 PTA DILATATION CATHETER
Doing adverse event research by hand?
The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.
Free to try · every answer cites its records
Report
- Report Number
- 9610978-2009-00096
- Event Type
- Malfunction
- Date Received
- April 30, 2009
- Date of Event
- June 11, 2007
- Report Date
- June 12, 2007
- Manufacturer
- CORDIS EUROPA, N.V.
- Product Code
- LIT
- PMA / PMN Number
- K032737
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
DURING AN INTERNAL REVIEW, IT WAS NOTED THAT THIS EVENT WAS BELIEVED TO HAVE BEEN SENT VIA ALTERNATIVE SUMMARY REPORTING (ASR). HOWEVER, IT WAS NOT SUBMITTED ON THE ASR REPORT AS IT DID NOT MEET THE GUIDELINES FOR ASR REPORTING AT THE CURRENT TIME THE REPORT WAS SUBMITTED (JULY 2007). THEREFORE, A 3500A WAS GENERATED TO REPORT THIS EVENT. DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) TO THE SUPERFICIAL FEMORAL ARTERY (SFA), THE BALLOON WAS REPORTED TO HAVE RUPTURED. THERE WAS NO REPORTED PT INJURY. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND SHOWED THAT THIS LOT OF PRODUCTS MET ALL REQUIREMENTS PER THE APPLICABLE MANUFACTURING QUALITY PLAN (MQP). THIS REVIEW WAS EXTENDED TO SUBASSEMBLY (B) (4) WITH LOT NUMBER 0101070562. THIS REVIEW CONFIRMED THAT SUBASSEMBLIES MET ALL REQUIREMENTS PER THE APPLICABLE MQP AS WELL, INCLUDING BURST TEST VALUES. WITH THE LIMITED AMOUNT OF INFORMATION AVAILABLE AND WITHOUT THE RETURN OF THE PRODUCT OR FILMS OF THE PROCEDURE, IT IS NOT POSSIBLE TO DETERMINE THE RELATIONSHIP BETWEEN THE DEVICE AND THE EVENT. HOWEVER, VESSEL CHARACTERISTICS MAY HAVE CONTRIBUTED TO THE REPORTED EVENT.
THE BALLOON RUPTURED AT 5 ATMOSPHERES OF PRESSURE. THE LESION BEING TREATED WAS THE SUPERFICIAL FEMORAL ARTERY. ONCE THE POWERFLEX P3 ((B) (4)) BALLOON BURST, CONTRAST MEDIA WAS OBSERVED LEAKING FROM THE BALLOON. ANOTHER BALLOON (SAME SIZE) WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO INJURY TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POWERFLEX P3 PTA DILATATION CATHETER | PTA CATHETERS (LIT) | LIT | CORDIS EUROPA, N.V. | NA | R0207045 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PROCEDURE.| A .035 GLIDEWIRE WAS USED DURING THE INDEX |