MOTO PARTIAL KNEE TIBIAL TRAY FIX CEMENTED S1 RM
Report
- Report Number
- 3005180920-2022-00407
- Event Type
- Injury
- Date Received
- May 31, 2022
- Date of Event
- May 2, 2022
- Report Date
- July 14, 2022
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- HSX
- UDI-DI
- 07630030896705
- PMA / PMN Number
- K162084
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 03 MAY 2022: LOT 182120: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11-JULY-2018. EXPIRATION DATE: 2023-06-20. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DEPARTMENT. FEW MONTHS AFTER PRIMARY UKA IN A LARGELY OVERWEIGHT PATIENT, THE TIBIAL BONE GIVES WAY AND THE BASEPLATE SUBSIDES. SUBSTITUTION TO A TOTAL KNEE BECOMES NECESSARY. THIS INCONVENIENCE WAS NOT RELATED TO A DEVICE DEFECT OR MALFUNCTION.
DEVICES NOT AVAILABLE ANYMORE, AS REPORTED BY THE SALES AGENT ON 14 JUNE 2022.
REVISION SURGERY AT 7 MONTHS FROM PRIMARY FOR TIBIAL IMPLANT SUBSIDENCE. THE SUBSIDENCE IS LIKELY DUE TO PATIENT WEIGHT AND INSUFFICIENT BONE STABILITY. ALL IMPLANTS WERE REVISED AND GMK SPHERE WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1599052 | MOTO PARTIAL KNEE TIBIAL TRAY FIX CEMENTED S1 RM | KNEE UNICOMPARTIMENTAL TIBIAL TRAY CEMENTED | HSX | MEDACTA INTERNATIONAL SA | 02.18.TF1.RM | 182120 | 07630030896705 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Female | Required Intervention |