FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 14522623 · Received May 27, 2022

Report

Report Number
3008642652-2022-05299
Event Type
Malfunction
Date Received
May 27, 2022
Date of Event
May 2, 2022
Report Date
May 26, 2022
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005005
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER UP) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO A DEFECTIVE U104 PXA PROCESSOR ON THE COMPUTER/ANALOG BOARD. THE ROOT CAUSE FOR THE DEFECTIVE U104 PXA PROCESSOR COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DAMAGED MONITOR.

Description of Event or Problem · 0

A US DISTRIBUTOR REPORTED THAT A MONITOR WILL NOT POWER UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320648 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 00855778005005

Patients

Seq Age Sex Outcome Treatment
1 Unknown