FDA Adverse Event Malfunction Summary report: N

HMS PLUS INSTRUMENT

MDR report key: 14520932 · Received May 27, 2022

Report

Report Number
2184009-2022-00123
Event Type
Malfunction
Date Received
May 27, 2022
Date of Event
April 29, 2022
Report Date
September 1, 2022
Manufacturer
PERFUSION SYSTEMS
Product Code
JOX
PMA / PMN Number
K101271
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT ANALYSIS: THE REPORTED NOT DISPENSING ISSUE WAS VERIFIED DURING SERVICE. UPON FUNCTIONAL INSPECTION TECHNICIAN FOUND DISPENSER SLIPPAGE. THE ISSUE WAS RESOLVED BY REPLACING THE SVC KIT SYRINGE MOTOR (M936399A001), WHEEL SYRINGE DRIVE (60606) AND THE TOGGLE ASSY SYRINGE DRIVE (60631). PREVENTIVE MAINTENANCE WAS PERFORMED PER SPECIFICATION. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CONCLUSION: COMPLAINT CONFIRMED FOR THE HMS PLUS INSTRUMENT GIVING INCORRECT READINGS DUE TO NOT DISPENSING. UPON FUNCTIONAL INSPECTION THE SERVICE TECHNICIAN FOUND DISPENSER SLIPPAGE. THE ISSUE WAS RESOLVED BY REPLACING THE SYRINGE MOTOR KIT, SYRINGE DRIVE WHEEL, AND SYRINGE DRIVE ASSEMBLY TOGGLE. NO PATIENT/CLINICAL SAFETY ISSUES REPORTED. TRENDS FOR ISSUES WITH THIS PRODUCT ARE REVIEWED AT QUARTERLY QUALITY MEETINGS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CORRECTION H4, DEVICE MANUFACTURE DATE: SEP-2002. MONTH AND YEAR ARE AVAILABLE ONLY (EXACT DAY UNKNOWN). MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A HMS PLUS INSTRUMENT, IT WAS REPORTED THAT IT WAS GIVING INCORRECT READINGS DUE TO NOT DISPENSING. NO ERROR CODES OR MESSAGES RECORDED. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1709631 HMS PLUS INSTRUMENT ANALYZER, HEPARIN, AUTOMATED JOX PERFUSION SYSTEMS 30514

Patients

Seq Age Sex Outcome Treatment
1 Unknown