SUPREME¿ ELECTROPHYSIOLOGY CATHETER, QUADRIPOLAR, 120 CM LENGTH, JSN¿ CURVE, ELE
Report
- Report Number
- 3008452825-2022-00442
- Event Type
- Malfunction
- Date Received
- May 26, 2022
- Date of Event
- May 18, 2022
- Report Date
- October 26, 2022
- Manufacturer
- ST. JUDE MEDICAL
- Product Code
- DRF
- UDI-DI
- 05414734201421
- PMA / PMN Number
- K002976
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED ENTANGLEMENT REMAINS UNKNOWN.
RELATED MANUFACTURING REF: 3008452825-2022-00441. DURING THE PROCEDURE, THE CATHETERS BECAME ENTANGLED WITH EACH OTHER. WHILE MAPPING IN THE RIGHT ATRIUM, THE HD GRID CATHETER WENT INTO THE RIGHT VENTRICLE (RV) AND THE RV CATHETER (JSN) WENT THROUGH THE OPENING BETWEEN THE SPLINES OF THE HD GRID CATHETER. THE PHYSICIAN WAS UNAWARE THAT THE RV CATHETER WENT THROUGH THE SPLINES AND PULLED THE HD GRID CATHETER BACK INTO THE SRO SHEATH TO REMOVE IT. WHEN DOING SO, THE JSN CATHETER WAS PULLED INTO THE SHEATH AND PULLED HARDER TO REMOVE HD GRID. FLUOROSCOPY REVEALED THE JSN CATHETER WAS DOUBLED OVER AND STICKING OUT OF THE TOP OF THE SRO SHEATH. BOTH CATHETERS WERE STUCK IN THE SHEATH. TECH SERVICES WAS CONTACTED AND WHILE ON THE PHONE, THE PHYSICIAN PUSHED AND THE TWO CATHETERS FINALLY CAME OUT THE DISTAL END OF THE SRO AND WAS ABLE TO REMOVE THE JSN, THEN THE GRID. BOTH CATHETERS WERE INTACT, BUT BENT. THERE WAS NO HARM TO THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1851871 | SUPREME¿ ELECTROPHYSIOLOGY CATHETER, QUADRIPOLAR, 120 CM LENGTH, JSN¿ CURVE, ELE | CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING | DRF | ST. JUDE MEDICAL | 401450 | 8215465 | 05414734201421 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | ADVISOR¿ HD GRID CATHETER, SE. |