BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML
Report
- Report Number
- 1119779-2022-00766
- Event Type
- Malfunction
- Date Received
- May 26, 2022
- Date of Event
- April 27, 2022
- Report Date
- August 5, 2022
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- MDB
- UDI-DI
- 30382902451229
- PMA / PMN Number
- K974883
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6 INVESTIGATION SUMMARY: MATERIAL 245122 IS MANUFACTURED BY REHYDRATING THE MEDIA COMPONENTS WITH USP PURIFIED WATER, AND THOROUGHLY MIXING UNTIL A HOMOGENEOUS SOLUTION IS OBTAINED. THE TUBES ARE FILLED, CAPPED, TORQUED, AND THEN LABELED BY MACHINE PER STANDARD OPERATING PROCEDURE (SOP). THE TUBES ARE TERMINALLY AUTOCLAVED IN AN AIR OVER PRESSURE (AOP) AUTOCLAVE, PER MANUFACTURING INSTRUCTIONS, USING A VALIDATED CYCLE. POST AUTOCLAVING, TUBES ARE PACKAGED INTO FINAL SHIPPING CONFIGURATIONS. THE BATCH HISTORY RECORD REVIEW FOR BATCH 1230008 WAS SATISFACTORY PER INTERNAL PROCEDURES. FORMULATION, FILLING, AND AUTOCLAVING PROCESSES WERE WITHIN SPECIFICATIONS. QC INSPECTION AND TESTING WERE SATISFACTORY AT TIME OF RELEASE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON THIS BATCH. RETENTION SAMPLES FROM BATCH 1230008 (100 TUBES) WERE AVAILABLE FOR INSPECTION. THE RETENTION SAMPLES SHOWED NO MEDIA DEFECTS IN 100/100 TUBES FROM VISUAL INSPECTION. THE RETENTION SAMPLES WERE TESTED FOR GROWTH AND ANTIBIOTIC SUSCEPTIBILITY. ALL PERFORMANCE TESTING WAS SATISFACTORY PER PROCEDURES FOR GROWTH AND ANTIBIOTIC SUSCEPTIBILITY AND THE NEGATIVE CONTROL REMAINED NEGATIVE THROUGHOUT TESTING. TWO PHOTOS WERE RECEIVED TO ASSIST WITH THE INVESTIGATION: THE FIRST PHOTO SHOWS MICROGRAPH RESULTS ON THE COMPUTER NO PRODUCT INFORMATION CAN BE SEEN FROM THE PHOTO. THE SECOND PHOTO ALSO SHOWS MICROGRAPH RESULTS ON THE COMPUTER. NO PRODUCT INFORMATION CAN BE SEEN IN THE PHOTO. WITHOUT PRODUCT INFORMATION A PHOTO ALONE CANNOT CONFIRM A COMPLAINT. A PHOTO MUST PROVIDE THE PRODUCT, THE PRODUCT DEFECT, AND IMPORTANT PRODUCT INFORMATION SUCH AS BATCH/LOT NUMBER. NO RETURNS WERE RECEIVED TO ASSIST WITH THE INVESTIGATION. THIS COMPLAINT CANNOT BE CONFIRMED. BD WILL CONTINUE TO TREND COMPLAINTS FOR PERFORMANCE. RISK MANAGEMENT REVIEW INDICATES THE POTENTIAL RISK OF THE DEFECT REPORTED WAS ASSESSED AS SEVERITY S4, PER BALTRMLMBACTECMGITAPH, REV 03, ID 1.11. QN: (B)(4). DW: (B)(4).
IT WAS REPORTED WHILE TESTING WITH BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML FALSE NEGATIVE RESULTS WERE OBTAINED. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT, AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE A POSITIVE MGIT CULTURE TUBE WHICH CAME OFF THE MGIT AS NEGATIVE AT 6 WEEKS. UPON ZN MICROSCOPY WE NOTE A VERY SCANTY BACTERIAL LOAD. WE WERE EXPECTING THE LOAD TO BE MUCH GREATER, AS IT WAS CULTURED IN DUPLICATE AND THE OTHER TUBE CAME OFF POSITIVE AFTER ONLY 3 WEEKS WITH A HIGH BACTERIAL LOAD.
IT WAS REPORTED WHILE TESTING WITH BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML FALSE NEGATIVE RESULTS WERE OBTAINED. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT, AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE A POSITIVE MGIT CULTURE TUBE WHICH CAME OFF THE MGIT AS NEGATIVE AT 6 WEEKS. UPON ZN MICROSCOPY WE NOTE A VERY SCANTY BACTERIAL LOAD. WE WERE EXPECTING THE LOAD TO BE MUCH GREATER, AS IT WAS CULTURED IN DUPLICATE AND THE OTHER TUBE CAME OFF POSITIVE AFTER ONLY 3 WEEKS WITH A HIGH BACTERIAL LOAD.