DAVINCI XI
Report
- Report Number
- 2955842-2022-11810
- Event Type
- Malfunction
- Date Received
- May 26, 2022
- Date of Event
- April 26, 2022
- Report Date
- April 26, 2022
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110720
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE REPORTED ISSUE COULD NOT BE REPLICATED DURING FIELD EVALUATION. FSE TESTED THE SYSTEM AND CONFIRMED NO ISSUE WITH ANY OF THE USM ARMS AND THE SYSTEM. AS A PREVENTIVE MEASURE, FSE PERFORMED ELECTRICAL AND MECHANICAL ADJUSTMENTS ON USM4. FOLLOWING THIS, SYSTEM TESTED AND VERIFIED AS READY FOR USE. A REVIEW OF THE SITE'S COMPLAINT HISTORY DOES NOT SHOW ANY ADDITIONAL COMPLAINTS RELATED TO THIS PRODUCT AND/OR THIS EVENT. NO IMAGE OR PROCEDURE VIDEO WAS PROVIDED FOR REVIEW. VERIFICATION VIA RFE LOGS CONFIRMS THE OCCURRENCE OF A PROCEDURE ON THE REPORTED EVENT DATE OF (B)(6) 2022 USING SYSTEM SK2511 MATCHING THE DOCUMENTED EVENT DETAILS. NO FURTHER REVIEW IS REQUIRED AS THE LOGS WERE REVIEWED/ANALYZED AS PART OF THE ISI TSE AND ISI FSE'S INVESTIGATION BASED ON THE INFORMATION AVAILABLE AT THIS TIME, THIS COMPLAINT IS BEING CLASSIFIED AS A REPORTABLE EVENT DUE TO THE FOLLOWING CONCLUSION: IT WAS ALLEGED THAT A UNIVERSAL SURGICAL MANIPULATOR (USM) MOVED WITH UNINTUITIVE MOTION (E.G. THE INSTRUMENT UNEXPECTEDLY JERKED/JUMPED/SWUNG/BOWED, MOVED IN AN UNEXPECTED/UNINTENDED/UNKNOWN WAY). UNINTUITIVE MOTION COULD LEAD TO SUBSEQUENT TISSUE DAMAGE. WHILE THERE WAS NO HARM OR INJURY TO THE PATIENT, THE REPORTED FAILURE MODE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF IT WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED CHOLECYSTECTOMY SURGICAL PROCEDURE, THE ROBOTICS COORDINATOR (ROCO) CALLED AN INTUITIVE SURGICAL INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) TO REPORT THAT UNIVERSAL SURGICAL MANIPULATOR (USM4) MOVES UNITUITIVELY WHEN GRIPPING A VESSEL. USER STATED THAT THE ISSUE WITH USM4 HAPPENS INTERMITTENTLY. TSE VIEWED SYSTEM LOGS WITH NO ERRORS NOTED ON ANY OF THE USM ARMS OR THE SYSTEM. SURGEON PROCEEDED WITH THE CASE. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1711542 | DAVINCI XI | PATIENT SIDE CART, 4-ARM | NAY | INTUITIVE SURGICAL, INC | 380652-44 | N/A | 00886874110720 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |