FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 14517378 · Received May 26, 2022

Report

Report Number
2955842-2022-11810
Event Type
Malfunction
Date Received
May 26, 2022
Date of Event
April 26, 2022
Report Date
April 26, 2022
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110720
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE REPORTED ISSUE COULD NOT BE REPLICATED DURING FIELD EVALUATION. FSE TESTED THE SYSTEM AND CONFIRMED NO ISSUE WITH ANY OF THE USM ARMS AND THE SYSTEM. AS A PREVENTIVE MEASURE, FSE PERFORMED ELECTRICAL AND MECHANICAL ADJUSTMENTS ON USM4. FOLLOWING THIS, SYSTEM TESTED AND VERIFIED AS READY FOR USE. A REVIEW OF THE SITE'S COMPLAINT HISTORY DOES NOT SHOW ANY ADDITIONAL COMPLAINTS RELATED TO THIS PRODUCT AND/OR THIS EVENT. NO IMAGE OR PROCEDURE VIDEO WAS PROVIDED FOR REVIEW. VERIFICATION VIA RFE LOGS CONFIRMS THE OCCURRENCE OF A PROCEDURE ON THE REPORTED EVENT DATE OF (B)(6) 2022 USING SYSTEM SK2511 MATCHING THE DOCUMENTED EVENT DETAILS. NO FURTHER REVIEW IS REQUIRED AS THE LOGS WERE REVIEWED/ANALYZED AS PART OF THE ISI TSE AND ISI FSE'S INVESTIGATION BASED ON THE INFORMATION AVAILABLE AT THIS TIME, THIS COMPLAINT IS BEING CLASSIFIED AS A REPORTABLE EVENT DUE TO THE FOLLOWING CONCLUSION: IT WAS ALLEGED THAT A UNIVERSAL SURGICAL MANIPULATOR (USM) MOVED WITH UNINTUITIVE MOTION (E.G. THE INSTRUMENT UNEXPECTEDLY JERKED/JUMPED/SWUNG/BOWED, MOVED IN AN UNEXPECTED/UNINTENDED/UNKNOWN WAY). UNINTUITIVE MOTION COULD LEAD TO SUBSEQUENT TISSUE DAMAGE. WHILE THERE WAS NO HARM OR INJURY TO THE PATIENT, THE REPORTED FAILURE MODE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF IT WERE TO RECUR.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED CHOLECYSTECTOMY SURGICAL PROCEDURE, THE ROBOTICS COORDINATOR (ROCO) CALLED AN INTUITIVE SURGICAL INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) TO REPORT THAT UNIVERSAL SURGICAL MANIPULATOR (USM4) MOVES UNITUITIVELY WHEN GRIPPING A VESSEL. USER STATED THAT THE ISSUE WITH USM4 HAPPENS INTERMITTENTLY. TSE VIEWED SYSTEM LOGS WITH NO ERRORS NOTED ON ANY OF THE USM ARMS OR THE SYSTEM. SURGEON PROCEEDED WITH THE CASE. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1711542 DAVINCI XI PATIENT SIDE CART, 4-ARM NAY INTUITIVE SURGICAL, INC 380652-44 N/A 00886874110720

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES