FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB

MDR report key: 14509930 · Received May 26, 2022

Report

Report Number
2032227-2022-195244
Event Type
Malfunction
Date Received
May 26, 2022
Date of Event
February 1, 2012
Report Date
May 25, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
0000006139949043
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THIS RECORD WAS CONVERTED. FOR ADDITIONAL INFORMATION ABOUT THE ANALYSIS, PLEASE SEE SAP. UNIT HAD MOTOR ERROR ON REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO VERIFY E-70 ALARM IN HISTORY SCREEN DUE TO MOTOR ERROR DURING REWINDING. HD (B)(6) 2012.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER WAS UNABLE TO SUCCESSFULLY CLEAR THE ALARM. CUSTOMER PERFORMED SELF TEST AND THE TEST PASSED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE TO USE THE PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2451036 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB A5723NABJ 0000006139949043

Patients

Seq Age Sex Outcome Treatment
1 62 YR Male