FDA Adverse Event Injury Summary report: N

BIGFOOT UNITY DIABETES MANAGEMENT SYSTEM

MDR report key: 14504418 · Received May 26, 2022

Report

Report Number
3016525500-2022-00005
Event Type
Injury
Date Received
May 26, 2022
Date of Event
April 23, 2022
Report Date
May 24, 2022
Manufacturer
BIGFOOT BIOMEDICAL, INC.
Product Code
QLG
UDI-DI
00850003506227
PMA / PMN Number
K202145
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER NARRATIVE: NO DEVICE ISSUES WERE REPORTED BY THE CUSTOMER. BIGFOOT CONDUCTED A THOROUGH INVESTIGATION AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. DATA LOGS FROM THE PATIENT'S UNITY SYSTEM WERE REVIEWED AND SHOWED THAT THE SYSTEM ISSUED A LOW GLUCOSE ALERT PER SPECIFICATION AND THE ALERT WAS ACKNOWLEDGED. AFTER INITIAL ACKNOWLEDGEMENT, THE SYSTEM ISSUED ALERTS AGAIN WHEN THE CUSTOMER'S GLUCOSE REACHED THE LOW AND VERY LOW LEVELS AND THESE ALERTS RESOLVED PER SPECIFICATION. IF BIGFOOT LEARNS OF ANY NEW INFORMATION IN RELATION TO THIS CASE, ANOTHER INVESTIGATION WILL BE PERFORMED, AND A FOLLOW-UP REPORT WILL BE SUBMITTED. ALL PERTINENT INFORMATION AVAILABLE TO BIGFOOT HAS BEEN SUBMITTED.

Description of Event or Problem · 0

A CUSTOMER EXPERIENCED SYMPTOMS OF SEVERE HYPOGLYCEMIA INCLUDING LIGHTHEADEDNESS, STUMBLING/FALLING, FEELING LETHARGIC, AND HAVING A HARD TIME REMEMBERING. THE CUSTOMER REQUIRED ASSISTANCE FROM THEIR STEP-SON WHO PROVIDED SOME CARBOHYDRATES (HAMBURGER STEAK, SALAD, AND SLAW) TO TREAT THE LOW GLUCOSE CONDITION. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342857 BIGFOOT UNITY DIABETES MANAGEMENT SYSTEM DIABETES MANAGEMENT SYSTEM QLG BIGFOOT BIOMEDICAL, INC. FG-300185 2590 00850003506227

Patients

Seq Age Sex Outcome Treatment
1 72 YR Female Other