FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 14504330 · Received May 26, 2022

Report

Report Number
1000113657-2022-00303
Event Type
Malfunction
Date Received
May 26, 2022
Date of Event
April 27, 2022
Report Date
July 13, 2022
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON (B)(6) 2022 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED THE REPLACEMENT PRODUCTS HAD NOT RESOLVED INITIAL CONCERN - REFERENCE INTERNAL REPORT REFERENCE NUMBER (B)(4).

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 13-JUL-2022: D9: DEVICE AVAILABLE FOR EVALUATION. H3: DEVICE EVALUATED BY MANUFACTURER. H6: UPDATED FDA¿S TYPE, FINDINGS AND CONCLUSIONS CODES. H10: METER AND TEST STRIPS WERE RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND NO DEFECT FOUND. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 213, 250, 206, 236, 240MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 130-160MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULT OF 286MG/DL USING TRUE METRIX METER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 10/20/2022 AND TEST STRIPS WERE OPENED 10 DAYS PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 213 MG/DL DATE: (B)(6) 2021 TIME: 5:30 AM FASTING. RESULT 2: 250 MG/DL DATE: (B)(6) 2021 TIME: AM FASTING. RESULT 3: 206 MG/DL DATE: (B)(6) 2022 TIME: AM FASTING. RESULT 4: 236 MG/DL DATE: (B)(6) 2022 TIME: AM FASTING. RESULT 5: 240 MG/DL DATE: (B)(6) 2022 TIME: AM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331531 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL ZY4317S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown