FDA Adverse Event Malfunction Summary report: N

EM 2400, MAIN MODULE

MDR report key: 14501803 · Received May 25, 2022

Report

Report Number
1416980-2022-02646
Event Type
Malfunction
Date Received
May 25, 2022
Date of Event
April 29, 2022
Report Date
June 9, 2022
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
NEP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CORRECTION MADE TO G1: BAXTER HEALTHCARE - ENGLEWOOD 14445 GRASSLANDS DR ENGLEWOOD CO. 80112 (PREVIOUSLY SUBMITTED AS PLEXUS CORPORATION 2500 MILLBROOK DRIVE BUFFALO GROVE, IL. 60089). H10: THE ACTUAL DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION WAS PERFORMED AND THE TOP PLATE WAS OBSERVED SCRATCHED, THE MAIN MODULE COVER WAS STAINED AND HAD CORRODED PAINT, AND A SCRATCHED PUMP COVER AND WORN RUBBER FEET WAS OBSERVED. UNKNOWN FLUIDS WERE FOUND ON THE UNIT. FUNCTIONAL TESTS INCLUDING ELECTRICAL SAFETY AND FUNCTIONAL SYSTEM LEVEL TESTS WERE PERFORMED AND NO ISSUES WERE NOTED. DURING INSPECTION OF THE POWER SUPPLY, BURNT TRACES WERE IDENTIFIED ON THE POWER SUPPLY BOARD (PCB) AROUND COMPONENTS U4, U8 AND CAPACITOR C27 FROM FLUID INGRESS. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE BURNING ODOR WAS DUE TO THE DAMAGED POWER SUPPLY BOARD, HOWEVER THE CAUSE OF THE DAMAGE COULD NOT BE DETERMINED. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT EM 2400, MAIN MODULE AUTOMATED COMPOUNDING DEVICE (ACD) SMELLED LIKE BURNING. THIS ISSUE WAS IDENTIFIED DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391883 EM 2400, MAIN MODULE SYSTEM/DEVICE, PHARMACY COMPOUNDING NEP BAXTER HEALTHCARE CORPORATION NA NA

Patients

Seq Age Sex Outcome Treatment
1 Unknown