FDA Adverse Event Injury Summary report: N

DRX-REVOLUTION MOBILE X-RAY SYSTEM (PLUS)

MDR report key: 14501337 · Received May 25, 2022

Report

Report Number
1317307-2022-00007
Event Type
Injury
Date Received
May 25, 2022
Date of Event
March 23, 2022
Report Date
May 24, 2022
Manufacturer
CARESTREAM HEALTH INC.
Product Code
IZL
UDI-DI
60889978618897
PMA / PMN Number
K191025
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CARESTREAM HEALTH HAS EVALUATED THE DEVICE. THE INVESTIGATION FOUND THAT THERE WERE NO DEVICE MALFUNCTIONS AND THE INCIDENT WAS CAUSED BY THE USER NOT FOLLOWING THE INSTRUCTIONS FOR USE (IFU). THE SITE CONTACT REPORTED CONFIRMED THAT THE TECH DID NOT USE BOTH HANDS, ONE HAND HOLDING OPEN A DOOR, TO DRIVE THE REVOLUTION AS SPECIFIED IN THE IFU. THERE WAS NO DEVICE MALFUNCTION. THIS WAS DUE TO USER ERROR. THE SERVICE PERSONNEL COMMUNICATED TO THE OPERATORS TO USE BOTH HANDS, PER THE INSTRUCTIONS FOR USE (IFU), WHILE DRIVING THE SYSTEM. THE SITE HAS ALSO STARTED USING RUBBER DOOR STOPS TO HOLD OPEN DOORS WHEN DRIVING THE DRX REVOLUTION THROUGH DOORWAYS TO MITIGATE THESE TYPES OF INCIDENTS. CARESTREAM HAS COMPLETED THIS INVESTIGATION.

Description of Event or Problem · 0

ON (B)(6) 2022, CARESTREAM HEALTH (CSH) WAS INFORMED, VIA EMAIL, OF AN INCIDENT RELATED TO THE DRX-REVOLUTION PLUS MOBILE XRAY SYSTEM WHICH OCCURRED AT DR (B)(6) ON (B)(6) 2022. PER CARESTREAM HEALTH SERVICE PERSONNEL INVESTIGATION WHICH CONFIRMED THE FOLLOWING: THE SYSTEM IN "AS FOUND" CONDITION: NORMAL WORKING CONDITIONS WITH NO ISSUES. AT THE TIME OF THE INCIDENT, THE SYSTEM WAS BEING DRIVEN BY A SITE TECH. PER THE REPORT, THE TECH/OPERATOR WAS HOLDING A DOOR OPEN WITH ONE HAND AND ATTEMPTING TO DRIVE THE DRX REVOLUTION PLUS SYSTEM WITH THE OTHER HAND WHEN BACKING INTO THE INTENSIVE CARE UNIT AND DROVE THE SYSTEM OVER HER FOOT. THE TECH WAS ABLE TO DRIVE THE SYSTEM FORWARD OFF OF HER FOOT. THE INCIDENT RESULTED IN A TOE INJURY WHERE THE TECH LOST THE TOE NAIL. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2093963 DRX-REVOLUTION MOBILE X-RAY SYSTEM (PLUS) DRX-REVOLUTION MOBILE X-RAY SYSTEM (PLUS) IZL CARESTREAM HEALTH INC. 60889978618897

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other