SPECTRUM IQ INFUSION PUMP
Report
- Report Number
- 1314492-2022-02088
- Event Type
- Injury
- Date Received
- May 25, 2022
- Date of Event
- April 22, 2022
- Report Date
- June 21, 2022
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- UDI-DI
- 00085412610900
- PMA / PMN Number
- K220417
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
G4 PMA/510K # OR BLA #: K220417. THE DEVICE WAS RECEIVED FOR EVALUATION. DURING FUNCTIONAL TESTING, 'INFUSION PROBLEM' WAS UNABLE TO BE REPRODUCED. EVALUATION PERFORMED FLOW ACCURACY TESTING WITH PASSING RESULTS WITH ERRORS OF 0.43% AND 0.90% WHICH ARE WITHIN SPECIFICATION. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. A DEVICE HISTORY REVIEW REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE REPORTED CONDITION WAS NOT VERIFIED. NO CORRECTION IS REQUIRED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SPECTRUM IQ PUMP HAD A FLOW ISSUE DURING THE INFUSION OF CITRATE (50 ML/H) RESULTING IN A MARKED DECREASE IN BICARBONATE WHICH LED TO METABOLIC ACIDOSIS AND HAEMODYNAMIC INSTABILITY WITH HIGH FLOW-RATE AMINE REQUIREMENTS. CONTINUOUS VENO-VENOUS HAEMOFILTRATION THERAPY WITH CITRATE AND CALCIUM INFUSIONS WERE STOPPED. THE PATIENT WAS TREATED WITH CALCIUM CHLORIDE SOLUTION AND BICARBONATE SOLUTION (STAT ). PATIENT OUTCOME WAS NOT REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2094957 | SPECTRUM IQ INFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE CORPORATION | NA | NA | 00085412610900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention | CITRATE| SET JC8386 |