SPECTRUM IQ INFUSION PUMP

MDR report key: 14499473 · Received May 25, 2022

Report

Report Number
1314492-2022-02088
Event Type
Injury
Date Received
May 25, 2022
Date of Event
April 22, 2022
Report Date
June 21, 2022
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
UDI-DI
00085412610900
PMA / PMN Number
K220417
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

G4 PMA/510K # OR BLA #: K220417. THE DEVICE WAS RECEIVED FOR EVALUATION. DURING FUNCTIONAL TESTING, 'INFUSION PROBLEM' WAS UNABLE TO BE REPRODUCED. EVALUATION PERFORMED FLOW ACCURACY TESTING WITH PASSING RESULTS WITH ERRORS OF 0.43% AND 0.90% WHICH ARE WITHIN SPECIFICATION. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. A DEVICE HISTORY REVIEW REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE REPORTED CONDITION WAS NOT VERIFIED. NO CORRECTION IS REQUIRED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SPECTRUM IQ PUMP HAD A FLOW ISSUE DURING THE INFUSION OF CITRATE (50 ML/H) RESULTING IN A MARKED DECREASE IN BICARBONATE WHICH LED TO METABOLIC ACIDOSIS AND HAEMODYNAMIC INSTABILITY WITH HIGH FLOW-RATE AMINE REQUIREMENTS. CONTINUOUS VENO-VENOUS HAEMOFILTRATION THERAPY WITH CITRATE AND CALCIUM INFUSIONS WERE STOPPED. THE PATIENT WAS TREATED WITH CALCIUM CHLORIDE SOLUTION AND BICARBONATE SOLUTION (STAT ). PATIENT OUTCOME WAS NOT REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2094957 SPECTRUM IQ INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION NA NA 00085412610900

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention CITRATE| SET JC8386