MID-C 105L
Report
- Report Number
- 3013461531-2022-00019
- Event Type
- Malfunction
- Date Received
- May 25, 2022
- Report Date
- May 25, 2022
- Manufacturer
- APIFIX LTD
- Product Code
- QGP
- PMA / PMN Number
- H170001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION: A REVIEW OF THE DHR DEMONSTRATED THAT THE MID-C SYSTEM WAS MANUFACTURED, TESTED, AND RELEASED ACCORDING TO APIFIX SPECIFICATIONS SURGEON X-RAY INFORMATION ANALYSIS: THE PATIENT WAS INITIALLY OPERATED ON (B)(6) 2018. THE 3 YEARS FOLLOW-UP RAISED A SUSPICION OF RATCHET MECHANISM FAILURE/BREAKAGE. THE 3-YEARS X-RAY INDICATED A SMALL FRACTURE FRAGMENT INSIDE THE RATCHET MECHANISM AREA. IN ADDITION, THE X-RAYS INDICATED A RATCHET MALFUNCTION (RESULTING IN A BACKUP OF THE DISTRACTION). RATCHET MALFUNCTION CAN RESULT FROM PHYSICAL TRAUMA, PRACTICING HIGH-DEMAND SPORTS, AND TISSUE GROWTH INSIDE THE RATCHET MECHANISM. RATCHET MALFUNCTION MAY BE A COINCIDENTAL FINDING AND MAY ALSO BE REPORTED TOGETHER WITH PAIN/CURVE PROGRESSION/NOISE. RISK ASSESSMENT: REOPERATION EVENTS ARE A KNOWN RISK THAT WAS ASSESSED AND RECORDED BY THE PRODUCT RISK ASSESSMENT DMS-777 REV Q THIS COMPLAINT DOES NOT CHANGE THE OCCURRENCES RATE. THE TOTAL RATE OF RATCHET MALFUNCTION FOR ANY REASON POSTS ECO- 59 IS 1.23%. THE COMPANY CONTINUES TO MONITOR THE EFFECTIVENESS OF THE MITIGATION.
PATIENT (B)(6) WAS OPERATED ON (B)(6) 2018. THE SURGEON REPORTED THAT THE THREE-YEAR FOLLOW-UP X-RAY INDICATED A FRACTURE OF THE INTERNAL RATCHET MECHANISM. THE PATIENT HAS A LOSS OF CORRECTION AND NO PAIN. NO ACCIDENT OCCURRED. A REVISION IS PLANNED, AND FUSION IS AN OPTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2114822 | MID-C 105L | POSTERIOR RATCHETING ROD SYSTEM | QGP | APIFIX LTD | MID-C 105L | AF 11-04-17 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |