FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 14494179 · Received May 25, 2022

Report

Report Number
9610877-2022-53478
Event Type
Malfunction
Date Received
May 25, 2022
Date of Event
May 10, 2022
Report Date
May 25, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K163614. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE PRISM CRACKED. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE PHYSICAL DAMAGE APPLIED ON THE OBJECTIVE PRISM. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(OBJECTIVE LENS BROKEN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2267747 PENTAX VIDEO DUODENOSCOPE - I10 STANDARD FDT HOYA CORPORATION PENTAX TOKYO OFFICE ED34-I10T

Patients

Seq Age Sex Outcome Treatment
1 Unknown