FDA Adverse Event
Other
Summary report: N
ICELOCK CLUTCH
MDR report key: 1448855
·
Received August 14, 2009
Report
- Report Number
- 1836248-2009-00002
- Event Type
- Other
- Date Received
- August 14, 2009
- Date of Event
- June 17, 2009
- Report Date
- July 10, 2009
- Manufacturer
- OSSUR H/F
- Product Code
- ISS
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
CONCLUSIONS: THE REPORTED COUNTER CLOCKWISE SPIN CAN OCCUR IF THE CLUTCH MECHANISM IS NOT TORQUED TO THE FORCE SPECIFIED IN THE INSTRUCTIONS FOR USE.
Description of Event or Problem · 1
"PT WALKED 8 BLOCKS TO GYM. WHILE ENTERING THE GYM AND GOING THROUGH THE DOOR, HE CAME OUT OF HIS LEG AND FELL ONTO HIS PROSTHESIS AND DISLOCATED HIS SHOULDER. HE WAS IN THE HOSPITAL FOR 10 DAYS"
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ICELOCK CLUTCH | ISS | OSSUR H/F | L-211000 | HF081111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Hospitalization |