Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED ¿SUTURE BROKE WHILE ADVANCING THE KNOT¿ COULD NOT BE CONFIRMED AS THE EXACT CONDITIONS ENCOUNTERED BY THE DEVICE DURING THE PROCEDURE COULD NOT BE REPLICATED IN THE TEST ENVIRONMENT AND THE ENTIRE SUTURE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE MANUFACTURING RECORDS REVEALED NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE INVESTIGATION WAS UNABLE TO DETERMINE A CONCLUSIVE CAUSE FOR THE REPORTED DIFFICULTIES; HOWEVER, THE TREATMENT APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.D4 - LOT NO UPDATED FROM 1110841 TO 2021241. D4 - EXPIRATION DATE UPDATED FROM 9/30/2023 TO 12/31/2023. H4 - DEVICE MFG DATE UPDATED FROM 11/8/2021 TO 2/12/2022.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER A PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY INTERVENTIONAL PROCEDURE USING A 8F SHEATH. REPORTEDLY, THE SUTURE BROKE WHILE ADVANCING THE KNOT. ANOTHER PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.