PLATE MUELLER HINTON II AGAR 20 EA
Report
- Report Number
- 1119779-2022-00746
- Event Type
- Malfunction
- Date Received
- May 21, 2022
- Date of Event
- April 26, 2022
- Report Date
- May 24, 2022
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- JTZ
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
OEM MANUFACTURE: THE MANUFACTURING LOCATION FOR THIS PRODUCT IS (B)(4). THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD DS HEADQUARTERS IN (B)(4) HAS BEEN LISTED, AND THE SPARKS FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US BUT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD BBL¿ MUELLER HINTON II AGAR CATALOG NUMBER 211438 WHICH IS A PREAMENDMENT DEVICE.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVAL YES, D9: RETURNED TO MANUFACTURER ON: 06-MAY-2022. INVESTIGATION SUMMARY: RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR EVALUATION/TESTING AND UPON COMPLETION, NO ISSUES WERE OBSERVED RELATING CONTAMINATION. WE CONFIRMED THE ISSUE AS A REPORT FROM RETURNED SAMPLE. PHOTO PROVIDED SHOWS CONTAMINATION AS WELL. THE ISSUE WAS CONTAMINATION. NO COMPLAINT HISTORY. NO ISSUE IN DEVICE HISTORY RECORD REVIEW OR RETENTION SAMPLES. NO TREND.
IT WAS REPORTED THAT WHILE USING PLATE MUELLER HINTON II AGAR 20 EA BIOLOGICAL CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ACCORDING TO THE CUSTOMER'S REPORT, THE MEDIA WAS FOUND TO BE CONTAMINATED."
IT WAS REPORTED THAT WHILE USING PLATE MUELLER HINTON II AGAR 20 EA BIOLOGICAL CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ACCORDING TO THE CUSTOMER'S REPORT, THE MEDIA WAS FOUND TO BE CONTAMINATED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 583435 | PLATE MUELLER HINTON II AGAR 20 EA | CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH | JTZ | BECTON, DICKINSON & CO. (SPARKS) | 2039414 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |