FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE¿ INSULIN SYRINGE

MDR report key: 14465232 · Received May 21, 2022

Report

Report Number
1920898-2022-00357
Event Type
Malfunction
Date Received
May 21, 2022
Date of Event
May 9, 2022
Report Date
June 7, 2022
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: CUSTOMER RETURNED (1) LOOSE 1/2CC, 12.7MM SYRINGE. CUSTOMER STATES THAT THE BARREL IS DAMAGED. THE RETURNED SYRINGE WAS EXAMINED AND EXHIBITED A PIECE OF THE BARREL BROKEN OFF RANGING FROM THE 0-10 UNIT MARKINGS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1242322. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ INSULIN SYRINGE BARREL WAS DAMAGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: "ACCORDING TO THE CUSTOMER'S REPORT, UPON UNPACKING, THE HCP NOTICED THAT THERE WAS A HOLE ON THE BARREL."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ INSULIN SYRINGE BARREL WAS DAMAGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: "ACCORDING TO THE CUSTOMER'S REPORT, UPON UNPACKING, THE HCP NOTICED THAT THERE WAS A HOLE ON THE BARREL.".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1368216 BD ULTRA-FINE¿ INSULIN SYRINGE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 1242322

Patients

Seq Age Sex Outcome Treatment
1 Unknown