XEN 45 GTS (AUSTRALIA)
Report
- Report Number
- 3011299751-2022-00087
- Event Type
- Injury
- Date Received
- May 20, 2022
- Report Date
- November 8, 2022
- Manufacturer
- ALLERGAN (IRVINE)
- Product Code
- KYF
- PMA / PMN Number
- K161457
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE EVENT OF "KERATITIS" IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. THE REPORTER HAS DECLINED TO PROVIDE ALLERGAN FURTHER INFORMATION REGARDING EVENT, PRODUCT, AND PATIENT DETAILS.
ARTICLE CITATION: ONG, J., MUKHTAR, A., YU, D., & MORGAN, W. H. "OUTCOMES OF 45 M GELATIN STENT SURGERY OVER 24-MONTH FOLLOW-UP." CLINICAL & EXPERIMENTAL OPHTHALMOLOGY. 2022 OCT 26. HTTPS://DOI.ORG/10.1111/CEO.14181.
HEALTHCARE PROFESSIONAL REPORTED AN UNKNOWN NUMBER OF XEN®45 GTS EVENTS IN UNKNOWN EYE FROM AN UNPUBLISHED LITERATURE ARTICLE DESCRIBED AS "HYPHEMA, DELLEN, HYPOTONY WITH SHALLOW ANTERIOR CHAMBER, HYPOTONY VISCOELASTIC FILL, AND CHOROIDAL EFFUSIONS."
THROUGH AN UPDATED UNPUBLISHED ARTICLE, "THE OUTCOMES OF 45 UM GELATIN STENT SURGERY OVER 24 MONTH FOLLOW-UP," FURTHER REPORTED 111 EYES REQUIRED AT LEAST ONE BLEB NEEDLING, 221 EYES REQUIRED MEDICATION TO REDUCE INTRAOCULAR PRESSURE (IOP), 41 EYES EXPERIENCED HYPHEMA, 8 EYES EXPERIENCED DELLEN, 16 EYES EXPERIENCED HYPOTONY WITH SHALLOW ANTERIOR CHAMBER, 1 EYE EXPERIENCED HYPOTONY WITH VISCOELASTIC FILL, AND 10 EYES EXPERIENCED CHOROIDAL EFFUSIONS.
ADDITIONAL INFORMATION PROVIDED THE PUBLISHED ARTICLE, "THE OUTCOMES OF 45 UM GELATIN STENT SURGERY OVER 24 MONTH FOLLOW-UP."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 563566 | XEN 45 GTS (AUSTRALIA) | IMPLANT, EYE VALVE | KYF | ALLERGAN (IRVINE) | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Female |