FDA Adverse Event Malfunction Summary report: N

BIGFOOT UNITY DIABETES MANAGEMENT SYSTEM

MDR report key: 14464292 · Received May 20, 2022

Report

Report Number
3016525500-2022-00004
Event Type
Malfunction
Date Received
May 20, 2022
Date of Event
April 21, 2022
Report Date
May 19, 2022
Manufacturer
BIGFOOT BIOMEDICAL, INC.
Product Code
QLG
PMA / PMN Number
K202145
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER NARRATIVE: NO COMPLAINTS OR ADVERSE EVENTS WERE REPORTED FROM THE FIELD. BIGFOOT IS REPORTING THIS ISSUE PROACTIVELY SINCE THERE IS A REMOTE POSSIBILITY OF MISSING CRITICAL ALERTS. THIS IS AN ISSUE INTRODUCED BY APPLE IOS 15 AND LATER.

Description of Event or Problem · 0

A BIGFOOT EMPLOYEE DISCOVERED THE FOLLOWING SCENARIO: THERE EXISTS A MULTIPLE-FAULT USE CASE FOR ALL IOS 15 VERSIONS THAT IF THE APP IS IN THE FOREGROUND FOR A PERIOD OF GREATER THAN FIVE MINUTES OR MORE (E.G. DISABLED LOCK SCREEN), THE MODAL FOR NOTIFICATIONS WILL SHOW VISUALLY TO THE USER BUT DUE TO THE ISSUE WITH APPLE IOS 15, THE SOUND AND VIBRATION WILL NOT ANNUNCIATE. THIS CASE IS AN ISSUE WITH APPLE IOS 15 VERSIONS AND NOT SPECIFIC TO THE BIGFOOT UNITY APP, WHERE THE OPERATING SYSTEM DOES NOT SEND VIBRATION/SOUND IN THE CASE THAT A SOUND/VIBE-ONLY NOTIFICATION IS TRIGGERED WHEN THE APP IS IN THE FOREGROUND. THIS WOULD MANIFEST IF THE USER WERE TO DISABLE THE LOCK OR SLEEP FEATURE ON THEIR PHONE. THERE IS A REMOTE POSSIBILITY THAT THIS MALFUNCTION, IF IT WERE TO RECUR, COULD RESULT IN MISSING A LOW GLUCOSE ALERT IF A PATIENT IS NOT LOOKING AT THEIR PHONE. NO COMPLAINTS FROM THE FIELD HAVE BEEN RECEIVED AS OF (B)(6) 2022.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1658770 BIGFOOT UNITY DIABETES MANAGEMENT SYSTEM DIABETES MANAGEMENT SYSTEM QLG BIGFOOT BIOMEDICAL, INC. SW-300842, V1.4.5

Patients

Seq Age Sex Outcome Treatment
1 Unknown