BD LUER-LOK¿ SYRINGE
Report
- Report Number
- 9610847-2022-00189
- Event Type
- Malfunction
- Date Received
- May 20, 2022
- Date of Event
- April 13, 2022
- Report Date
- June 23, 2022
- Manufacturer
- BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
- Product Code
- FMF
- UDI-DI
- 00382903097029
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTED AND/OR ADDITIONAL INFORMATION. MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: D.4. MEDICAL DEVICE LOT #: 2021835. D.4. MEDICAL DEVICE EXPIRATION DATE: 30-SEP-2026. H.4. DEVICE MANUFACTURE DATE: 16-FEB-2022. D.4. MEDICAL DEVICE LOT #: 2021832. D.4. MEDICAL DEVICE EXPIRATION DATE: 30-SEP-2026. H.4. DEVICE MANUFACTURE DATE: 03-FEB-2022. D.4. MEDICAL DEVICE LOT #: 2028783. D.4. MEDICAL DEVICE EXPIRATION DATE: 31-OCT-2026. H.4. DEVICE MANUFACTURE DATE: 22-FEB-2022.
H6: INVESTIGATION SUMMARY: OUR QUALITY ENGINEER INSPECTED THE 3 PHOTOS AND 24 LOOSE SAMPLES SUBMITTED FOR EVALUATION. THE REPORTED ISSUES OF FOREIGN MATTER, WERE CONFIRMED UPON INSPECTION AND TESTING OF THE SAMPLES. SEVEN SAMPLES DISPLAYED EXCESSIVE SILICONE INSIDE THE SYRINGE. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND WITHIN THE BD LUER-LOK¿ SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INTERNAL PARTICULATE.
IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND WITHIN THE BD LUER-LOK¿ SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INTERNAL PARTICULATE.
IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND WITHIN THE BD LUER-LOK¿ SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INTERNAL PARTICULATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1200054 | BD LUER-LOK¿ SYRINGE | PISTON SYRINGE | FMF | BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. | SEE H.10. | 00382903097029 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |