JAMSHIDI (TJ) NEEDLE BM 8G X 4 ASP MAC - TJM4008
Report
- Report Number
- 1625685-2022-00056
- Event Type
- Malfunction
- Date Received
- May 20, 2022
- Date of Event
- May 9, 2022
- Report Date
- July 5, 2022
- Manufacturer
- CAREFUSION, INC
- Product Code
- KNW
- UDI-DI
- 10885403044564
- PMA / PMN Number
- K913306
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PR (B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. DEVICE PROBLEM CODE: (B)(4). PATIENT PROBLEM CODE: (B)(4).
(B)(4) FOLLOW-UP EMDR FOR DEVICE EVALUATION: ONE PHOTO AND ONE SAMPLE WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IT IS OBSERVED A SLIVER APPROXIMATELY 3 QUARTERS OF AN INCH IN LENGTH WITH A WIDTH APPROXIMATELY THE SAME AS A PENCIL LEAD. THE SMALL SLIVER WAS EXAMINED UNDER MAGNIFICATION. THE PHYSICAL CHARACTERISTICS OF THE ITEM INDICATED IT WAS PARTIALLY METAL OF 2 OR MORE COLORS AND MAGNETIC. THEREFORE, THE REPORTED FAILURE MODE WAS CONFIRMED. A DEVICE HISTORY REVIEW COULD NOT BE COMPLETED AS NO BATCH NUMBER WAS PROVIDED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO IDENTIFY A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND DEFECT WILL CONTINUE TO BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS. H3 OTHER TEXT : SEE MANUFACTURE NARRATION.
IRON FILINGS FOUND IN THE TROCAR (SEALED IN THE BAG FOR VISUALIZATION) TO OUR KNOWLEDGE, THE PATIENT WAS NOT INVOLVED.
IRON FILINGS FOUND IN THE TROCAR (SEALED IN THE BAG FOR VISUALIZATION). TO OUR KNOWLEDGE, THE PATIENT WAS NOT INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1001235 | JAMSHIDI (TJ) NEEDLE BM 8G X 4 ASP MAC - TJM4008 | INSTRUMENT, BIOPSY | KNW | CAREFUSION, INC | TJM4008 | UNKNOWN | 10885403044564 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |