FDA Adverse Event
Death
Summary report: N
HOSPITAL BED AND HILL ROM AIR MATTRESS
MDR report key: 1445133
·
Received August 19, 2009
Report
- Report Number
- 2182305-2009-00029
- Event Type
- Death
- Date Received
- August 19, 2009
- Date of Event
- July 18, 2009
- Report Date
- August 14, 2009
- Manufacturer
- JOERNS HEALTHCARE INC. FOR BED
- Product Code
- FNL
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED TO MANUFACTURER, BY FACILITY, PER FACILITY IN THE EARLY MORNING, THEY FOUND A FEMALE RESIDENT EXPIRED. CLAIMING THAT THE BED WAS IN ITS LOW POSITION, HER LOWER BODY WAS LYING ON THE FLOOR WITH HER NECK AND HEAD AGAINST SIDERAIL. PRELIMINARY AUTOPSY FINDINGS: PRIMARY CAUSE OF DEATH ASSOCIATED WITH ASPHYXIATION ENTRAPMENT, SECONDARY HEART DISEASE. ALL EQUIPMENT WAS IN WORKING ORDER. FACILITY WAS USING A HILL ROM PW-50 AIR MATTRESS WITH OUR U795 BED AND F17TM 1/2 RAILS. AS DESCRIBED THIS WOULD BE CONSIDERED AS A ZONE 3: BETWEEN THE RAIL AND THE MATTRESS ENTRAPMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOSPITAL BED AND HILL ROM AIR MATTRESS | BED, HOSPITAL ELECTRIC & HILL ROM AIR MATTRESS | FNL | JOERNS HEALTHCARE INC. FOR BED | U795 W/ F17TM 1/2 RAILS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 93 YR | Death |