FDA Adverse Event Death Summary report: N

HOSPITAL BED AND HILL ROM AIR MATTRESS

MDR report key: 1445133 · Received August 19, 2009

Report

Report Number
2182305-2009-00029
Event Type
Death
Date Received
August 19, 2009
Date of Event
July 18, 2009
Report Date
August 14, 2009
Manufacturer
JOERNS HEALTHCARE INC. FOR BED
Product Code
FNL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED TO MANUFACTURER, BY FACILITY, PER FACILITY IN THE EARLY MORNING, THEY FOUND A FEMALE RESIDENT EXPIRED. CLAIMING THAT THE BED WAS IN ITS LOW POSITION, HER LOWER BODY WAS LYING ON THE FLOOR WITH HER NECK AND HEAD AGAINST SIDERAIL. PRELIMINARY AUTOPSY FINDINGS: PRIMARY CAUSE OF DEATH ASSOCIATED WITH ASPHYXIATION ENTRAPMENT, SECONDARY HEART DISEASE. ALL EQUIPMENT WAS IN WORKING ORDER. FACILITY WAS USING A HILL ROM PW-50 AIR MATTRESS WITH OUR U795 BED AND F17TM 1/2 RAILS. AS DESCRIBED THIS WOULD BE CONSIDERED AS A ZONE 3: BETWEEN THE RAIL AND THE MATTRESS ENTRAPMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOSPITAL BED AND HILL ROM AIR MATTRESS BED, HOSPITAL ELECTRIC & HILL ROM AIR MATTRESS FNL JOERNS HEALTHCARE INC. FOR BED U795 W/ F17TM 1/2 RAILS

Patients

Seq Age Sex Outcome Treatment
1 93 YR Death