FDA Adverse Event Malfunction Summary report: N

UNKNOWN

MDR report key: 14444905 · Received May 19, 2022

Report

Report Number
3002808486-2022-00671
Event Type
Malfunction
Date Received
May 19, 2022
Report Date
May 18, 2022
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

CUSTOMER (PERSON) NOT PROVIDED, INFORMATION PROVIDED BY MCSWEENEY LANGEVIN LLC. OCCUPATION: NON-HEALTHCARE PROFESSIONAL. INVESTIGATION: FILTER TILT HAS BEEN REPORTED. POTENTIAL CAUSES MAY INCLUDE FILTER PLACEMENT IN IVCS WITH DIAMETERS LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER); AND (OR) A FAILED RETRIEVAL ATTEMPT. EXCESSIVE FILTER TILT MAY CONTRIBUTE TO DIFFICULT OR FAILED RETRIEVAL; VENA CAVA WALL PENETRATION/PERFORATION; AND (OR) RESULT IN LOSS OF FILTER EFFICIENCY. POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: UNACCEPTABLE FILTER TILT. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 0

IT IS ALLEGED THAT "PATIENT RECEIVED A COOK TULIP FILTER. PATIENT IS ALLEGING FILTER TILT. HOSPITAL AND MEDICAL RECORDS HAVE NOT BEEN PROVIDED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2408736 UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Male