FDA Adverse Event Injury Summary report: N

GUNTHER TULIP JUGULAR VENA CAVA FILTER SET

MDR report key: 14444563 · Received May 19, 2022

Report

Report Number
1820334-2022-00940
Event Type
Injury
Date Received
May 19, 2022
Report Date
May 28, 2024
Manufacturer
COOK INC
Product Code
DTK
UDI-DI
10827002330174
PMA / PMN Number
K072240
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. CUSTOMER (PERSON) NOT PROVIDED, INFORMATION PROVIDED BY ONDERLAW, LLC OCCUPATION: NON-HEALTHCARE PROFESSIONAL INVESTIGATION: FILTER INTERACTS WITH IVC WALL, E.G. PENETRATION/PERFORATION/EMBEDMENT. THIS MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: TRAUMA TO ADJACENT STRUCTURES, VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. RPN/LOT IS UNKNOWN, HOWEVER, THE DEVICE IS MANUFACTURED AND INSPECTED ACCORDING TO CURRENT CONTROLS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

INVESTIGATION: INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: MIGRATION, TILT, LEFT/RIGHT 'EYE STROKE', ABDOMINAL PAIN, DEPRESSION. FILTER OR FILTER FRAGMENT MIGRATION AND (OR) EMBOLIZATION (E.G., MOVEMENT TO THE HEART OR LUNGS) HAS BEEN REPORTED. FILTER OR FILTER FRAGMENT MOVEMENT HAS OCCURRED IN BOTH THE CRANIAL AND CAUDAL DIRECTION AND MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE FILTER PLACEMENT IN IVCS WITH DIAMETERS SMALLER OR LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; DEPLOYMENT INTO THROMBUS; DISLODGEMENT DUE TO LARGE THROMBUS BURDENS; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: FILTER MIGRATION, TRAUMA TO ADJACENT STRUCTURES. FILTER TILT HAS BEEN REPORTED. POTENTIAL CAUSES MAY INCLUDE FILTER PLACEMENT IN IVCS WITH DIAMETERS LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER); AND (OR) A FAILED RETRIEVAL ATTEMPT. UNKNOWN IF THE REPORTED LEFT/RIGHT 'EYE STROKE', ABDOMINAL PAIN, AND DEPRESSION ARE DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. CATALOG AND LOT NUMBERS ARE UNKNOWN, HOWEVER, THE ALLEGED TULIP IS MANUFACTURED AND INSPECTED ACCORDING TO CONTROLS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

IT IS ALLEGED THAT "PATIENT RECEIVED A COOK TULIP FILTER. PATIENT IS ALLEGING PERFORATION OF IVC. HOSPITAL AND MEDICAL RECORDS HAVE NOT BEEN PROVIDED."

Description of Event or Problem · 0

PATIENT ALLEGEDLY RECEIVED A GUNTHER TULIP INFERIOR VENA CAVA (IVC) FILTER ON (B)(6) 2012 VIA THE RIGHT INTERNAL JUGULAR VEIN DUE TO IVC THROMBOSIS. PATIENT IS ALLEGING DEVICE MIGRATION AND TILT, VENA CAVA PERFORATION. PATIENT NOTES AND FURTHER ALLEGES EXPERIENCING ABDOMINAL PAIN, LEFT 'EYE STROKE', RIGHT 'EYE STROKE', AND DEPRESSION. PER A REPORT FROM COMPUTED TOMOGRAPHY (CT): "CAVAL PERFORATION: YES. 4 O'CLOCK 3.9 MM GRADE 2. 8 O'CLOCK 5.2 MM GRADE 2. 10 O'CLOCK 9.0 MM GRADE 2. TILT: YES. ANTERIOR TILT MEASURING 11.50 DEGREES. MIGRATION: YES. APEX OF THE CAVAL FILTER IS AT THE LEVEL OF THE RENAL VEINS. PERTINENT NEGATIVES: NONE. ADDITIONAL FINDINGS: NONE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2319926 GUNTHER TULIP JUGULAR VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK COOK INC G33017 2874188 10827002330174

Patients

Seq Age Sex Outcome Treatment
1 NA Male Life Threatening