PACKAGE, RED 350P, 1X PAD-PAK-03, GERMAN, 350-STR-ST-10
Report
- Report Number
- 3004123209-2022-00075
- Event Type
- Malfunction
- Date Received
- May 19, 2022
- Date of Event
- May 3, 2022
- Report Date
- August 16, 2022
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- NSA
- PMA / PMN Number
- P160008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
HEARTSINE TECHNOLOGIES LTD HAS REQUESTED THE RETURN OF THE DEVICE FROM THE DISTRIBUTOR IN ORDER TO INVESTIGATE THE ALLEGED MALFUNCTION. UPON COMPLETION, THE CONCLUSIONS WILL BE SUBMITTED IN A FOLLOW-UP REPORT.
HEARTSINE'S INVESTIGATION OF THE DEVICE CONFIRMED THE REPORTED FAULT AS UPON RETURN THE DEVICE WAS UNABLE TO DELIVER A SHOCK. THE REPORTED FAULT WAS ATTRIBUTED TO A FAILED CAPACITOR, WHICH FORMS PART OF THE SWITCHING CIRCUITRY USED TO DELIVER PHASE ONE OF SHOCK DELIVERY.THE DEVICE WAS SCRAPPED BY HEARTSINE AND THE CUSTOMER WAS PROVIDED WITH A REPLACEMENT DEVICE. D2 UPDATE: DUE TO NEW FDA APPROVAL FOR THIS MODEL, THE PRODUCT CODE HAS BEEN UPDATED FROM MKJ TO NSA.
THE DISTRIBUTOR CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE DOES NOT GIVE A SHOCK DURING TESTING. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.
THE DISTRIBUTOR CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE DOES NOT GIVE A SHOCK DURING TESTING. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1848279 | PACKAGE, RED 350P, 1X PAD-PAK-03, GERMAN, 350-STR-ST-10 | AUTOMATED EXTERNAL DEFIBRILLATOR | NSA | HEARTSINE TECHNOLOGIES LTD | 350P | ||
| 392393 | PACKAGE, RED 350P, 1X PAD-PAK-03, GERMAN, 350-STR-ST-10 | AUTOMATED EXTERNAL DEFIBRILLATOR | NSA | HEARTSINE TECHNOLOGIES LTD | 350P |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |