FDA Adverse Event Malfunction Summary report: N

PACKAGE, RED 350P, 1X PAD-PAK-03, GERMAN, 350-STR-ST-10

MDR report key: 14436996 · Received May 19, 2022

Report

Report Number
3004123209-2022-00075
Event Type
Malfunction
Date Received
May 19, 2022
Date of Event
May 3, 2022
Report Date
August 16, 2022
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
NSA
PMA / PMN Number
P160008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

HEARTSINE TECHNOLOGIES LTD HAS REQUESTED THE RETURN OF THE DEVICE FROM THE DISTRIBUTOR IN ORDER TO INVESTIGATE THE ALLEGED MALFUNCTION. UPON COMPLETION, THE CONCLUSIONS WILL BE SUBMITTED IN A FOLLOW-UP REPORT.

Additional Manufacturer Narrative · 0

HEARTSINE'S INVESTIGATION OF THE DEVICE CONFIRMED THE REPORTED FAULT AS UPON RETURN THE DEVICE WAS UNABLE TO DELIVER A SHOCK. THE REPORTED FAULT WAS ATTRIBUTED TO A FAILED CAPACITOR, WHICH FORMS PART OF THE SWITCHING CIRCUITRY USED TO DELIVER PHASE ONE OF SHOCK DELIVERY.THE DEVICE WAS SCRAPPED BY HEARTSINE AND THE CUSTOMER WAS PROVIDED WITH A REPLACEMENT DEVICE. D2 UPDATE: DUE TO NEW FDA APPROVAL FOR THIS MODEL, THE PRODUCT CODE HAS BEEN UPDATED FROM MKJ TO NSA.

Description of Event or Problem · 0

THE DISTRIBUTOR CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE DOES NOT GIVE A SHOCK DURING TESTING. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

Description of Event or Problem · 0

THE DISTRIBUTOR CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE DOES NOT GIVE A SHOCK DURING TESTING. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1848279 PACKAGE, RED 350P, 1X PAD-PAK-03, GERMAN, 350-STR-ST-10 AUTOMATED EXTERNAL DEFIBRILLATOR NSA HEARTSINE TECHNOLOGIES LTD 350P
392393 PACKAGE, RED 350P, 1X PAD-PAK-03, GERMAN, 350-STR-ST-10 AUTOMATED EXTERNAL DEFIBRILLATOR NSA HEARTSINE TECHNOLOGIES LTD 350P

Patients

Seq Age Sex Outcome Treatment
1 Unknown