FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 14436927 · Received May 19, 2022

Report

Report Number
3013756811-2022-49681
Event Type
Malfunction
Date Received
May 19, 2022
Date of Event
April 25, 2022
Report Date
April 25, 2022
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613762
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE CUSTOMER WAS ABLE TO SUCCESSFULLY CHARGE THE PUMP USING A DIFFERENT WALL ADAPTER. CUSTOMER¿S BLOOD GLUCOSE VALUE WAS BETWEEN 150-155 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1975992 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 00850006613762

Patients

Seq Age Sex Outcome Treatment
1 62 YR Male INSULIN: HUMALOG