FDA Adverse Event Malfunction Summary report: N

WRO 300H

MDR report key: 14436562 · Received May 19, 2022

Report

Report Number
14436562
Event Type
Malfunction
Date Received
May 19, 2022
Date of Event
April 28, 2022
Report Date
May 9, 2022
Manufacturer
MAR COR PURIFICATION, INC.
Product Code
FIP
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

PATIENT WAS STARTED ON DIALYSIS IN THE ICU, SHORTLY AFTER STARTING THE RO STARTED TO ALARM, "CONDUCTIVITY ALARM". IT WOULD SILENCE AND THE PATIENT CONTINUED TO RUN WITHOUT AN ISSUE WHILE THE TECH BROUGHT UP A NEW RO AND THE PATIENT WAS RINSED BACK AND HER MACHINE WAS HOOKED UP TO THE NEW RO AND SHE WAS RE-STARTED ON NEW RO AND COMPLETED DIALYSIS WITHOUT FURTHER ISSUES. HTM REPLACED MEMBRANE, AND PUT BACK IN SERVICE. WILL NOT BE RETURNING DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1815082 WRO 300H SUBSYSTEM, WATER PURIFICATION FIP MAR COR PURIFICATION, INC. RO,WRO 300 H,115V~,60HZ

Patients

Seq Age Sex Outcome Treatment
1 Unknown