FDA Adverse Event
Malfunction
Summary report: N
WRO 300H
MDR report key: 14436562
·
Received May 19, 2022
Report
- Report Number
- 14436562
- Event Type
- Malfunction
- Date Received
- May 19, 2022
- Date of Event
- April 28, 2022
- Report Date
- May 9, 2022
- Manufacturer
- MAR COR PURIFICATION, INC.
- Product Code
- FIP
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Description of Event or Problem · 0
PATIENT WAS STARTED ON DIALYSIS IN THE ICU, SHORTLY AFTER STARTING THE RO STARTED TO ALARM, "CONDUCTIVITY ALARM". IT WOULD SILENCE AND THE PATIENT CONTINUED TO RUN WITHOUT AN ISSUE WHILE THE TECH BROUGHT UP A NEW RO AND THE PATIENT WAS RINSED BACK AND HER MACHINE WAS HOOKED UP TO THE NEW RO AND SHE WAS RE-STARTED ON NEW RO AND COMPLETED DIALYSIS WITHOUT FURTHER ISSUES. HTM REPLACED MEMBRANE, AND PUT BACK IN SERVICE. WILL NOT BE RETURNING DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1815082 | WRO 300H | SUBSYSTEM, WATER PURIFICATION | FIP | MAR COR PURIFICATION, INC. | RO,WRO 300 H,115V~,60HZ |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |