FDA Adverse Event Injury Summary report: N

INPEN MMT-105NNPKNA NOVO NORDISK PINK

MDR report key: 14433476 · Received May 19, 2022

Report

Report Number
3012822846-2022-00744
Event Type
Injury
Date Received
May 19, 2022
Date of Event
May 6, 2022
Report Date
July 11, 2022
Manufacturer
COMPANION MEDICAL INC
Product Code
FMF
UDI-DI
000010862088000368
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 0

SERIAL NUMBER: N/A, SOFTWARE VERSION: N/A, COLOR: PINK, BATTERY LIFE REMAINING: N/A. CUSTOMER REPORTS: INPEN NOT WORKING AS INTENDED. PER VISUAL INSPECTION: NO PHYSICAL DAMAGE TO CAP, CARTRIDGE HOLDER OR INPEN WAS NOTED. THE LEADSCREW WAS RECEIVED 1/4 IT'S TRAVEL, THE LEADSCREW WAS REWOUND PROPERLY. LEADSCREW NOT BENT, ADVANCED WHEN DOSAGE KNOB WAS PRESSED DIALING A DOSAGE AND RETRACTED APPROPRIATELY. NO RESISTANCE WAS OBSERVED WHEN DOSING WITHOUT A CARTRIDGE INSTALLED. THE SCREW ADVANCED EVERY TIME 30.0U WAS DIALED AND DOSED UNTIL THE SCREW REACHED MAX EXTENSION. NO LEADSCREW ANOMALY WAS NOTED. IN CONCLUSION: PER NETA GLASER SAN DIEGO INVESTIGATION: TESTED OK. INPEN RECEIVED WORKING AS DESIGN. NO MALFUNCTIONS NOTED DURING TESTING THAT COULD AFFECT INSULIN DELIVERY. THEREFORE, THE CUSTOMER COMPLAINT OF INPEN NOT WORKING AS INTENDED COULD NOT BE CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THEY REPEAT PRIME IF NO INSULIN EXITS WITH INITIAL PRIME STILL THE INSULIN WAS NOT EXIT. CUSTOMER HAD EXPERIENCED HIGH BLOOD GLUCOSE VALUE. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 415 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER WILL DISCONTINUE TO USE THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1975781 INPEN MMT-105NNPKNA NOVO NORDISK PINK SYRINGE, PISTON FMF COMPANION MEDICAL INC MMT-105NNPKNA B0124 000010862088000368

Patients

Seq Age Sex Outcome Treatment
1 71 YR Female Other