COBAS® EGFR MUTATION TEST V2
Report
- Report Number
- 2243471-2022-00478
- Event Type
- Malfunction
- Date Received
- May 18, 2022
- Date of Event
- April 11, 2022
- Report Date
- May 17, 2022
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
- Product Code
- OWD
- UDI-DI
- 00875197005448
- PMA / PMN Number
- P150047
- Removal / Correction Number
- 2243471-08-24-2021-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ROCHE RECEIVED COMPLAINTS FROM CUSTOMERS REPORTING THE GENERATION OF FALSE MUTATION DETECTED RESULTS FOR THE EXON 20 INSERTION (EX20INS) MUTATION WHEN USING THE COBAS® EGFR MUTATION TEST V2. DURING IN-HOUSE TESTING USING CUSTOMER-PROVIDED FFPET SAMPLES, AN EX20INS FALSE MUTATION DETECTED RESULT WAS REPRODUCED FOR ONE OUT OF 8 FFPET SAMPLES, WHICH WAS PROCESSED FOLLOWING THE VALIDATED SAMPLE PREPARATION METHOD FROM THE INSTRUCTIONS FOR USE. ALTHOUGH THE MAJORITY OF CASES REPORTED WERE FROM USERS USING FFPET SAMPLES, THE GENERATION OF FALSE MUTATION DETECTED EX20INS RESULTS WITH PLASMA SPECIMENS WAS REPORTED IN ONE CASE. A FALSE MUTATION DETECTED EX20INS RESULT COULD LEAD TO HARM UNDER SPECIFIC SCENARIOS. CONSIGNEES WERE NOTIFIED OF THE ISSUE WITH INSTRUCTION TO FOLLOW THE INSTRUCTIONS FOR USE FOR SAMPLE INPUT REQUIREMENTS. ADDITIONALLY, IF AN EX20INS MUTATION DETECTED RESULT IS GENERATED WITH THE COBAS® EGFR MUTATION TEST V2, CUSTOMERS MUST CONFIRM THE RESULT WITH ANOTHER METHOD (E.G., SEQUENCING OR OTHER PCR-BASED TESTS). (B)(4).
DURING REVIEW OF CUSTOMER PROVIDED DATA, ONE SAMPLE WAS IDENTIFIED TO HAVE MADE A POTENTIAL FALSE "MUTATION DETECTED" RESULT FOR EXON20INS. NO RETEST WAS PERFORMED FOR THIS SAMPLE. THERE IS NO ALLEGATION OF HARM OR SERIOUS INJURY DUE TO THE EXON20INS CALLS. AN INVESTIGATION WAS CONDUCTED TO EVALUATE THE CUSTOMER ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1711119 | COBAS® EGFR MUTATION TEST V2 | SOMATIC GENE MUTATION DETECTION SYSTEM | OWD | ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG | NA | H14612 | 00875197005448 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |