FDA Adverse Event Malfunction Summary report: N

CONSTELLATION ULTRAVIT PROBE

MDR report key: 14427279 · Received May 18, 2022

Report

Report Number
1644019-2022-00376
Event Type
Malfunction
Date Received
May 18, 2022
Date of Event
April 28, 2022
Report Date
August 16, 2022
Manufacturer
ALCON RESEARCH, LLC - HOUSTON
Product Code
MLZ
UDI-DI
00380657514380
PMA / PMN Number
K093305
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION PROVIDED IN H.6., AND H.10. A SAMPLE WAS NOT RECEIVED AT THE MANUFACTURING SITE FOR EVALUATION; THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED. A REVIEW OF THE DEVICE HISTORY RECORD TRACEABLE TO THE REPORTED LOT NUMBER, INDICATES THAT THE REPORTED PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO THE REPORTED PRODUCT¿S ACCEPTANCE CRITERIA. A SAMPLE WAS NOT RECEIVED AT THE MANUFACTURING SITE AND THE DEVICE HISTORY RECORD REVIEW OF THE LOT NUMBER PROVIDED INDICATED THE PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO THE PRODUCT¿S ACCEPTABLE CRITERIA, THEREFORE; THE ROOT CAUSE FOR THE CUSTOMER COMPLAINT ISSUE CANNOT BE DETERMINED. THE EXACT ROOT CAUSE FOR THIS COMPLAINT IS UNKNOWN; THEREFORE, SPECIFIC ACTION WITH REGARDS TO THIS COMPLAINT CANNOT BE TAKEN. ALL PROBES ARE 100% VISUALLY INSPECTED AND TESTED FOR ACTUATION, ASPIRATION, AND CUT DURING MANUFACTURING. COMPLAINTS ARE REVIEWED AND MONITORED AT REGULAR INTERVALS FOR ANY SIGNIFICANT ADVERSE TRENDS. NO ADVERSE TRENDS HAVE BEEN OBSERVED ASSOCIATED WITH THE REPORTED PRODUCT AND EVENT. NO FURTHER ACTIONS ARE REQUIRED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A NURSE REPORTED VITRECTOMY PROBE CANNOT CUT DURING A VITRECTOMY SURGERY. THE PRODUCT WAS REPLACED AND THE SURGERY WAS COMPLETED. THERE WAS NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2753518 CONSTELLATION ULTRAVIT PROBE VITRECTOMY, INSTRUMENT CUTTER MLZ ALCON RESEARCH, LLC - HOUSTON NA 2482537H 00380657514380

Patients

Seq Age Sex Outcome Treatment
1 Unknown