FDA Adverse Event Injury Summary report: N

IMPL TWIST MTX 3.75 MM 13 MM

MDR report key: 14425331 · Received May 17, 2022

Report

Report Number
0002023141-2022-01200
Event Type
Injury
Date Received
May 17, 2022
Date of Event
March 15, 2022
Report Date
October 18, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K013494
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER: K943604.

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBERS 61898439 & 60889213. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHRS. LOTS WERE INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORDS ST# (B)(4) (61898439) & ST# (B)(4) (60889213) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBERS 61898439 & 60889213 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO PERI-IMPLANTITIS. EDEMA AND INFLAMMATION WERE REPORTED AS SYMPTOMS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1851029 IMPL TWIST MTX 3.75 MM 13 MM DENTAL IMPLANT DZE ZIMMER DENTAL 60889213

Patients

Seq Age Sex Outcome Treatment
1 72 YR Male Required Intervention