FDA Adverse Event Malfunction Summary report: N

MICRO SCREW 1.2MM X 6MM

MDR report key: 14412055 · Received May 16, 2022

Report

Report Number
2027754-2022-00020
Event Type
Malfunction
Date Received
May 16, 2022
Date of Event
May 6, 2022
Report Date
June 3, 2022
Manufacturer
OSTEOMED, LLC
Product Code
DZL
UDI-DI
00845694000482
PMA / PMN Number
K924138
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE AS THE DEVICE HAS NOT BEEN RECEIVED FOR EVALUATION. IF THE DEVICE IS RETURNED, A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 0

THE LOT NUMBERS FOR THIS EVENT WERE REPORTED AS 1154827, 1157319, AND 1160324, SO THE DEVICE HISTORY RECORDS (DHRS) WERE REVIEWED FOR THESE LOT NUMBERS. THE REVIEW REVEALED ALL INSPECTIONS PASSED, AND NO ANOMALIES WERE NOTED. SIX (6) SCREWS WERE RETURNED FOR EVALUATION. ADDITIONALLY, FRAGMENTS OF BROKEN SCREWS WERE RECEIVED. THE SIX (6) SCREWS WERE INSPECTED BY QUALITY CONTROL (QC) PERSONNEL PER THE INSPECTION PLAN.THE RESULTS FOR EACH OF THE SIX (6) SCREWS WERE AS FOLLOWS: ONE SCREW WAS OBSERVED TO BE UNUSED AND MET ALL SPECIFICATIONS. THE SECOND AND THIRD SCREWS HAD CRUCIFORMS THAT SHOWED SIGNS OF USAGE/EXHIBITED WEAR AND THEREFORE DID NOT MEET SPECIFICATIONS. THE FOURTH, FIFTH, AND SIXTH SCREWS WERE ONLY ABLE TO MEASURED FOR THE LENGTH BUT WERE UNABLE TO BE MEASURED FOR ALL OTHER FEATURES DUE TO SIGNIFICANT DAMAGE TO THE CRUCIFORM, WHICH IS INDICATIVE OF MULTIPLE FAILED ATTEMPTS OF INSERTION. A TWO YEAR REVIEW OF COMPLAINTS INDICATED THERE HAVE BEEN TWO (2) COMPLAINTS FOR THIS PART FAMILY OF SCREWS; HOWEVER, THE COMPLAINT IN THIS REPORT IS THE ONLY COMPLAINT RECEIVED FOR "CANNOT BE INSERTED INTO BONE AND SCREW WAS BROKEN/NOT STRAIGHT". REVIEW OF THE INSTRUCTIONS FOR USE (IFU) REVEALED THE FOLLOWING GUIDANCE IS PROVIDED: USE OF AN UNDERSIZED PLATE OR SCREW IN AREAS OF HIGH FUNCTIONAL STRESSES MAY LEAD TO IMPLANT FRACTURE AND FAILURE. USE OF SCREWS IN HIGH DENSITY BONE MAY LEAD TO IMPLANT FRACTURE OR FAILURE UPON INSERTION. USE OF EXCESSIVE TORQUE DURING INSERTION OF SCREWS MAY LEAD TO IMPLANT FAILURE. ALL OSTEOMED PLATES, SCREWS, AND INSTRUMENTATION MAY BE REQUIRED FOR EACH SURGERY. FAILURE TO USE DEDICATED, UNIQUE OSTEOMED INSTRUMENTS FOR EVERY STEP OF THE IMPLANTATION TECHNIQUE MAY COMPROMISE THE INTEGRITY OF THE IMPLANTED DEVICE, LEADING TO PREMATURE DEVICE FAILURE AND SUBSEQUENT PATIENT INJURY. FAILED DEVICES MAY REQUIRE RE-OPERATION AND REMOVAL. CAREFULLY INSPECT THE OSTEOMED IMPLANTS PRIOR TO USE. INSPECT THE INSTRUMENTS BEFORE AND AFTER EACH PROCEDURE TO ASSURE THEY ARE IN PROPER OPERATING CONDITION. INSTRUMENTS WHICH ARE FAULTY, DAMAGED, OR SUSPECT SHOULD NOT BE USED. THEY SHOULD BE REPLACED OR SENT TO OSTEOMED. THE ROOT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED BASED ON THE INFORMATION PROVIDED AND THE INVESTIGATION PERFORMED. OUT OF THE SIX (6) DEVICES RETURNED, ONE (1) DEVICE WHICH EXHIBITED NO SIGNS OF WEAR/USAGE MET SPECIFICATION PER THE INSPECTION PLAN WHILE THE FIVE (5) REMAINING DEVICES DID NOT MEET SPECIFICATION FOR ONE (1) OR MORE FEATURES SINCE DAMAGE TO THE CRUCIFORM WAS NOTED, WHICH IS INDICATIVE OF MULTIPLE REPEATED/FAILED ATTEMPTS OF INSERTION; HOWEVER, DUE TO UNKNOWN PROCEDURAL CONDITIONS, THE CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A FACIAL TRAUMA PROCEDURE, THE SURGEON USED THE PART NUMBER 204-1206 SCREWS, AND ONE OF THE FOLLOWING SITUATIONS OCCURRED: 1) THE SCREW COULD NOT BE INSERTED INTO BONE, AND THEN, THE SCREW WAS BROKEN OR "WAS NOT STRAIGHT" OR 2) THE SCREW WAS INSERTED INTO BONE, AND THEN, THE SCREW WAS BROKEN. IT WAS REPORTED PREDRILLING WAS PERFORMED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT; HOWEVER, IT WAS REPORTED THIS ISSUE RESULTED IN DELAY IN PROCEDURE AND FRAGMENT OF THE BROKEN SCREW REMAINS IN THE PATIENT. IT IS UNKNOWN HOW MANY SCREWS BROKES BROKE/FRAGMENTS REMAIN IN PATIENT. THE SURGEON USED 1.2MM SCREW WITH ANOTHER LENGTH TO COMPLETE THE SURGERY. INFORMATION RECEIVED INDICATED 14 SCREWS FROM THREE DIFFERENT BATCH/LOT NUMBERS WERE INVOLVED. THIS REPORT IS RELATED TO THE FOLLOWING REPORT NUMBERS FOR THE OTHER SCREWS INVOLVED IN THIS EVENT: 2027754-2022-00012, 2027754-2022-00013, 2027754-2022-00014, 2027754-2022-00015, 2027754-2022-00016, 2027754-2022-00017, 2027754-2022-00018, 2027754-2022-00019, 2027754-2022-00021, 2027754-2022-00022, 2027754-2022-00023, 2027754-2022-00024, AND 2027754-2022-00025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1388794 MICRO SCREW 1.2MM X 6MM SCREW, FIXATION, INTRAOSSEOUS DZL OSTEOMED, LLC 204-1206 1160324 00845694000482

Patients

Seq Age Sex Outcome Treatment
1 44 YR Female