FDA Adverse Event Injury Summary report: N

SYNCARDIA FREEDOM DRIVER

MDR report key: 14408532 · Received May 16, 2022

Report

Report Number
3003761017-2022-00044
Event Type
Injury
Date Received
May 16, 2022
Date of Event
April 7, 2022
Report Date
April 5, 2024
Manufacturer
SYNCARDIA SYSTEMS, LLC
Product Code
LOZ
UDI-DI
00858000003121
PMA / PMN Number
P030011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FREEDOM DRIVER WAS RETURNED TO SYNCARDIA FOR EVALUATION. THE RESULTS WILL BE PROVIDED IN A FOLLOW-UP MDR.

Additional Manufacturer Narrative · 0

DEVICE HISTORY RECORD (DHR) REVIEW CONFIRMED THAT FREEDOM DRIVER S/N (B)(6) WAS SERVICED AND PASSED ALL FUNCTIONAL TESTING PRIOR TO BEING RELEASED TO FINISHED GOODS. ALARM HISTORY AND PATIENT DATA FILE REVIEW FOUND NO NEW ALARMS. VISUAL INSPECTION OF EXTERNAL COMPONENTS FOUND THE EXTERNAL POWER TO MAIN BOARD CABLE PORT DAMAGED. VISUAL INSPECTION OF INTERNAL COMPONENTS FOUND CRACK IN BOTTOM RIGHT ANCHOR BOSS OF THE FRONT DRIVER HOUSING WITH A PIN-NUT PROTRUDING. INSPECTION ALSO FOUND A SCUFF/BURN MARK TO THE SPEAKER TO LCD RIBBON CABLE AND SECONDARY MOTOR OUT OF PRIMARY ALIGNMENT. FREEDOM DRIVER PASSED ALL AREAS OF FUNCTIONAL TESTING FOR ACCEPTANCE AT INCOMING INSPECTION WITH CABLE HELD IN PLACE WITH TAPE UNTIL PART COULD BE SCRAPPED AFTER INVESTIGATION. ADDITIONAL TESTING INCLUDED A HYPERVOLEMIA TEST TO MIMIC FLUID OVERLOAD PER CUSTOMER REPORT. FILL VOLUME DROPPED, REPLICATING CUSTOMER REPORTED EXPERIENCE, HOWEVER, THIS IS A COMMON OCCURRENCE WITH THE FREEDOM DRIVER SOFTWARE WHERE THE FILL VOLUME READINGS DROP WHEN PATIENT'S FLUID INTAKE IS TOO HIGH. THIS IS A KNOWN ISSUE THAT IS CURRENTLY BEING ADDRESSED WITH UPDATED SOFTWARE VERSIONS BUT IS NOT CONSIDERED A DEVICE MALFUNCTION. FAILURE INVESTIGATION FOR THIS COMPLAINT CONFIRMED THE REPORTED ISSUE. THE CUSTOMER COMPLAINT WAS REPLICATED DURING TESTING; ROOT CAUSE OF THE DROP IN FILL VOLUMES WAS DUE TO THE CURRENT FREEDOM DRIVER SOFTWARE VERSION THAT DOES NOT ACCOMMODATE ACCURATE READINGS ON A FLUID OVERLOADED PATIENT. THIS IS NOT A DEVICE A MALFUNCTION BUT A KNOWN ISSUE THAT IS CURRENTLY BEING ADDRESSED WITH UPDATED SOFTWARE. FAILURE INVESTIGATION IDENTIFIED NO TEST FAILURES OR DAMAGE THAT COULD HAVE CONTRIBUTED TO THE COMPLAINT. EXTERNAL AND INTERNAL DAMAGE WAS COSMETIC AND DID NOT DISRUPT THE FUNCTIONALITY OF THE DRIVER. NO DEVICE MALFUNCTION WAS FOUND. PATIENT CONDITION WAS HYPERVOLEMIC AND THEREFORE WERE OPERATING THE DEVICE OUTSIDE OF NORMOTENSIVE REQUIREMENTS. PATIENT'S CONDITION WAS STABILIZED WITH HOSPITAL INTERVENTION AND SWITCHED TO A BACKUP COMPANION 2 DRIVER WITHOUT ANY ADDITIONAL REPORTED ADVERSE IMPACT. THIS ISSUE WILL BE MONITORED AND TRENDED AS PART OF THE CUSTOMER COMPLAINT PROCESS. SYNCARDIA HAS COMPLETED ITS EVALUATION AND IS CLOSING THIS FILE. IF NEW OR ADDITIONAL INFORMATION IS RECEIVED IN THE FUTURE, SYNCARDIA WILL FILE A FOLLOW-UP MDR.

Description of Event or Problem · 0

THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE PATIENT WAS PLACED ON THIS FREEDOM DRIVER ON (B)(6) 2022. THROUGHOUT HIS TIME AT HOME, THE PATIENT FELT WELL. HE CAME IN FOR A CLINIC VISIT ON (B)(6) 2022 AND HIS TOTAL BILIRUBIN AND CREATININE WAS MEASURED AND INDICATED AN INCREASE OVER BASELINE POINTING TO A POSSIBLE FLUID OVERLOAD CONDITION. A TRANSESOPHAGEAL ECHOCARDIOGRAM WAS PERFORMED WHICH SHOWED THAT HIS RIGHT ATRIUM WAS DILATED THUS CONFIRMING A FLUID OVERLOAD PATIENT CONDITION. THE PATIENT WAS ADMITTED, SWITCHED TO A COMPANION 2 DRIVER AND MEDICAL TREATMENT WAS ADMINISTERED TO DECREASE HIS FLUID OVERLOAD CONDITION. THE CUSTOMER ALSO REPORTED THE PATIENT'S FLUID STATUS RETURNED TO NORMAL RANGE AND HE WAS TRANSITIONED BACK TO A FREEDOM DRIVER. THE PATIENT WAS DISCHARGED HOME (B)(6) 2022 AND HAS SUBSEQUENTLY BEEN SEEN IN THE CLINIC AS PART OF HIS NORMAL CARE AND WAS DOING WELL. THE CUSTOMER ALSO REPORTED THAT THE FILL VOLUME NUMBERS WERE NOT THE SAME ON THE COMPANION 2 DRIVER AS COMPARED TO THE FREEDOM DRIVER WHEN SUPPORTING THE PATIENT. THE HOSPITAL DID NOT THINK THE FREEDOM DRIVER FILL VOLUMES WERE READING CORRECTLY.

Description of Event or Problem · 0

THE CUSTOMER, A (B)(6) HOSPITAL, REPORTED THAT THE PATIENT WAS PLACED ON THIS FREEDOM DRIVER ON (B)(6) 2022. THROUGHOUT HIS TIME AT HOME, THE PATIENT FELT WELL. HE CAME IN FOR A CLINIC VISIT ON (B)(6) 2022 AND HIS TOTAL BILIRUBIN AND CREATININE WAS MEASURED AND INDICATED AN INCREASE OVER BASELINE POINTING TO A POSSIBLE FLUID OVERLOAD CONDITION. A TRANSESOPHAGEAL ECHOCARDIOGRAM WAS PERFORMED WHICH SHOWED THAT HIS RIGHT ATRIUM WAS DILATED THUS CONFIRMING A FLUID OVERLOAD PATIENT CONDITION. THE PATIENT WAS ADMITTED, SWITCHED TO A COMPANION 2 DRIVER AND MEDICAL TREATMENT WAS ADMINISTERED TO DECREASE HIS FLUID OVERLOAD CONDITION. THE CUSTOMER ALSO REPORTED THE PATIENT'S FLUID STATUS RETURNED TO NORMAL RANGE AND HE WAS TRANSITIONED BACK TO A FREEDOM DRIVER. THE PATIENT WAS DISCHARGED HOME (B)(6) 2022 AND HAS SUBSEQUENTLY BEEN SEEN IN THE CLINIC AS PART OF HIS NORMAL CARE AND WAS DOING WELL. THE CUSTOMER ALSO REPORTED THAT THE FILL VOLUME NUMBERS WERE NOT THE SAME ON THE COMPANION 2 DRIVER AS COMPARED TO THE FREEDOM DRIVER WHEN SUPPORTING THE PATIENT. THE HOSPITAL DID NOT THINK THE FREEDOM DRIVER FILL VOLUMES WERE READING CORRECTLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2427962 SYNCARDIA FREEDOM DRIVER EXTERNAL PNEUMATIC DRIVER LOZ SYNCARDIA SYSTEMS, LLC 00858000003121

Patients

Seq Age Sex Outcome Treatment
1 22 YR Male Hospitalization| R