FDA Adverse Event Malfunction Summary report: N

BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE

MDR report key: 14407804 · Received May 16, 2022

Report

Report Number
9610847-2022-00183
Event Type
Malfunction
Date Received
May 16, 2022
Date of Event
April 25, 2022
Report Date
July 5, 2022
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FPA
UDI-DI
30382903851004
PMA / PMN Number
K013621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY OUR QUALITY ENGINEER INSPECTED THE PHOTOGRAPH SUBMITTED FOR EVALUATION. BD RECEIVED ONE PHOTOGRAPH, ONE WHICH DISPLAYED THE LABEL FOR A 0.16ML BD Q-SYTE FROM LOT NUMBER 1210005. THE PHOTO PROVIDED FOR THIS INCIDENT DID NOT PRESENT SUFFICIENT EVIDENCE TO IDENTIFY OR CONFIRM THE ALLEGED FAILURE OR TO ESTABLISH A DEFINITE ROOT CAUSE. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH. SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 5 BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICES LEAKED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6): "IN THE PAST WEEK WE HAVE HAD 4 OR 5 LEAKING Q SITES. AFTER REPLACING IT, NO MORE LEAKS."

Description of Event or Problem · 0

IT WAS REPORTED THAT 5 BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICES LEAKED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM DUTCH: "IN THE PAST WEEK WE HAVE HAD 4 OR 5 LEAKING Q SITES. AFTER REPLACING IT, NO MORE LEAKS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1745020 BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE INTRAVASCULAR ADMINISTRATION SET FPA BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 385100 1210005 30382903851004

Patients

Seq Age Sex Outcome Treatment
1 Unknown