FDA Adverse Event Malfunction Summary report: N

BD¿ BULK STERILE CONVENIENCE PAK TIP CAP SYRINGE

MDR report key: 14407521 · Received May 16, 2022

Report

Report Number
1213809-2022-00286
Event Type
Malfunction
Date Received
May 16, 2022
Date of Event
March 1, 2022
Report Date
June 8, 2022
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
KYW
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.9. DEVICE AVAILABLE FOR EVAL?: YES. D.9. RETURNED TO MANUFACTURER ON: 27-MAY-2022. H.6. INVESTIGATION: FIFTY SAMPLE AND TWO PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IT IS OBSERVED THAT ONE SAMPLE WITH THREE BROWNISH-BLACK AREAS WHICH IS IDENTIFIED AS EMBEDDED FOREIGN MATTER. POTENTIAL ROOT CAUSE FOR THE EMBEDDED FOREIGN MATTER DEFECT IS ASSOCIATED WITH THE MOLDING PROCESS. THE EMBEDDED FOREIGN MATTER IS MOST LIKELY DEGRADED PLASTIC. THIS OCCURS WHEN THE RESIN IS EXPOSED TO PROLONGED HIGH TEMPERATURES INSIDE THE MOLDING MACHINE, SUCH AS DURING START UP. THIS TYPE OF DEFECT IS COSMETIC AND DOES NOT POSE RISK TO THE CUSTOMER. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED BATCH NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT BROWN FOREIGN MATTER WAS FOUND IN THE BD¿ BULK STERILE CONVENIENCE PAK TIP CAP SYRINGES STERILE PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE CUSTOMER COMPLAINED THAT BROWNISH SPOTS WERE FOUND IN THE PRODUCT OF THE STERILE PACK."

Description of Event or Problem · 0

IT WAS REPORTED THAT BROWN FOREIGN MATTER WAS FOUND IN THE BD¿ BULK STERILE CONVENIENCE PAK TIP CAP SYRINGES STERILE PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE CUSTOMER COMPLAINED THAT BROWNISH SPOTS WERE FOUND IN THE PRODUCT OF THE STERILE PACK.".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1745005 BD¿ BULK STERILE CONVENIENCE PAK TIP CAP SYRINGE LIQUID MEDICATION DISPENSER KYW BECTON DICKINSON MEDICAL SYSTEMS 1200274

Patients

Seq Age Sex Outcome Treatment
1 Unknown