FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 14388250 · Received May 13, 2022

Report

Report Number
2647876-2022-00136
Event Type
Malfunction
Date Received
May 13, 2022
Date of Event
April 14, 2022
Report Date
June 27, 2022
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
K113558
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. (B)(6). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.

Additional Manufacturer Narrative · 0

D.4. UPDATED MEDICAL DEVICE LOT # FIELD FROM UNKNOWN TO SEE H10. MULTIPLE LOT NUMBERS ARE REPORTED TO BE INVOLVED: D.4. MEDICAL DEVICE LOT #: 2013086 D.4. MEDICAL DEVICE EXPIRATION DATE: 10/31/2022 H.4. DEVICE MANUFACTURE DATE: 01/13/2022 D.4. MEDICAL DEVICE LOT #: 2040773 D.4. MEDICAL DEVICE EXPIRATION DATE: 11/30/2022 H.4. DEVICE MANUFACTURE DATE: 02/09/2022 D.4. MEDICAL DEVICE LOT #: 1252583 D.4. MEDICAL DEVICE EXPIRATION DATE: 06/30/2022 H.4. DEVICE MANUFACTURE DATE: 09/09/2021 D.4. MEDICAL DEVICE LOT #: 2082639 D.4. MEDICAL DEVICE EXPIRATION DATE: 01/31/2023 H.4. DEVICE MANUFACTURE DATE: 03/23/2022 D.4. MEDICAL DEVICE LOT #: 1314592 D.4. MEDICAL DEVICE EXPIRATION DATE: 08/31/2022 H.4. DEVICE MANUFACTURE DATE: 11/10/2021 D.4. MEDICAL DEVICE LOT #: 1055914 D.4. MEDICAL DEVICE EXPIRATION DATE: 12/31/2021 H.4. DEVICE MANUFACTURE DATE: 02/24/2021. H.6 INVESTIGATION SUMMARY: CATALOG 442023 BATCH NO. 2013086, 2040773, 2082639 &1314592. CUSTOMER REPORTED A CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH BACTEC PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS. CATALOG 442023, BATCH NO. UNKNOWN. CUSTOMER REPORTED A CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS NOT ABLE TO PERFORM AN INVESTIGATION TO THE RETENTION SAMPLES SINCE BATCH NUMBER IS UNKNOWN. A COMPLAINT HISTORY REVIEW CANNOT BE CONDUCTED RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE SINCE BATCH NUMBER IS UNKNOWN. THE BATCH HISTORY RECORD COULD NOT BE REVIEWED AS THE LOT NUMBER IS UNKNOWN NONETHELESS BATCH HISTORY RECORDS ARE ALWAYS REVIEWED PRIOR TO PRODUCT RELEASE. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. COMPLAINT IS UNCONFIRMED. NO CORRECTIVES ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATION PROCESS. CATALOG 442023, BATCH NO. 1252583. CUSTOMER REPORTED A CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. RETENTION SAMPLES COULD NOT BE VISUALLY INSPECTED/ TESTED FOR REPORTED DEFECT. BATCH HISTORY RECORD REVIEW WAS USED FOR THIS INVESTIGATION. BATCH/SENSOR HISTORY RECORDS WERE REVIEWED, AND ALL TESTING WERE WITHIN SPECIFICATION FOR PRODUCT RELEASE. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON BATCH HISTORY RECORD REVIEW. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS. CATALOG 442023, BATCH NO. 1055914 EXPIRED. CUSTOMER REPORTED A CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. UPON FURTHER EVALUATION IT WAS NOTICED THAT COMPLAINT RECEIVED WAS FROM A PRODUCT ALREADY EXPIRED. NO INVESTIGATION CAN BE CONDUCTED TO THE RETENTION SAMPLES. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. THE BATCH HISTORY RECORD COULD NOT BE REVIEWED AS THE LOT IS EXPIRED, NONETHELESS BATCH HISTORY RECORDS ARE ALWAYS REVIEWED PRIOR TO PRODUCT RELEASE. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. THE SPECIMEN MUST BE COLLECTED USING STERILE TECHNIQUES TO REDUCE THE CHANCE OF CONTAMINATION. DO NOT USE CULTURE VIALS PAST THEIR EXPIRATION DATE. COMPLAINT IS UNCONFIRMED. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) THE CUSTOMER OBSERVED BIOLOGICAL CONTAMINATION OF THE MEDIA. THIS EVENT OCCURRED 26 TIME(S). THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED BY THE CUSTOMER THAT THERE IS A POSSIBLE CONTAMINATION.", "BOTTLE SEQUENCE NUMBERS FOR POSSIBLY AFFECTED 442023 BOTTLES. 26 TOTAL: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) THE CUSTOMER OBSERVED BIOLOGICAL CONTAMINATION OF THE MEDIA. THIS EVENT OCCURRED 26 TIME(S). THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED BY THE CUSTOMER THAT THERE IS A POSSIBLE CONTAMINATION.", "BOTTLE SEQUENCE NUMBERS FOR POSSIBLY AFFECTED 442023 BOTTLES. 26 TOTAL: 449271089378, 449277720522, 449290289379, 449290289547, 449290289520, 449290289518, 449289033176, 449289032919, 449289033288, 449289033252, 449295779460, 449295779710, 449289033387, 449299193326, 449299204403, 449256127124, 449256127488, 449256127400, 449256122721, 449256127430, 449256126016, 449267593664, 449256126016, 449256126136, 449256125671, 449262006545."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
667756 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 442023 SEE H10 00382904420239

Patients

Seq Age Sex Outcome Treatment
1 Unknown