FDA Adverse Event Malfunction Summary report: N

TRI-TOME PC PROTECTOR

MDR report key: 14380197 · Received May 12, 2022

Report

Report Number
1037905-2022-00221
Event Type
Malfunction
Date Received
May 12, 2022
Date of Event
April 1, 2022
Report Date
May 12, 2022
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
KNS
UDI-DI
10827002491974
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PMA/510(K) # K172665. INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE CUTTING WIRE SECURING COMPONENT LOCATED NEAR THE DISTAL END OF THE SPHINCTEROTOME HAS DISCONNECTED FROM THE CATHETER. DUE TO THE CATHETER AND SECURING COMPONENT DISCONNECTION, THE DISTAL END OF THE CUTTING WIRE IS NO LONGER CONNECTED TO THE SPHINCTEROTOME CATHETER AT THE DISTAL END. THE SECURING COMPONENT HAS A LONGER SECTION MEASURING 3 MM AND A SHORTER SECTION MEASURING 2 MM THEREFORE NO PART OF THE DEVICE IS MISSING. THE CUTTING WIRE ALSO EXHIBITS SLIGHT EVIDENCE OF A CAUTERY APPLICATION (BLACKENING OF THE CUTTING WIRE WAS NOTED). A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A VISUAL EXAMINATION OF THE RETURNED DEVICE CONFIRMED THE CUTTING WIRE SECURING COMPONENT HAS SEPARATED FROM THE CATHETER. A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER CAN OCCUR IF THE TIP OF THE SPHINCTEROTOME IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER: "DO NOT OVER FLEX OR BOW TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE OR CAUSE CUTTING WIRE TO BREAK." OTHER FACTORS THAT CAN CONTRIBUTE TO SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER: "UPON REMOVING DEVICE FROM PACKAGE, UNCOIL AND STRAIGHTEN SPHINCTEROTOME. CAREFULLY REMOVE PRECURVED STYLET FROM CANNULATING TIP." THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. THE FUNCTIONAL TEST INCLUDES BOWING THE SPHINCTEROTOME TO ENSURE THE DISTAL END RESPONDS TO HANDLE MANIPULATION. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK TRI-TOME PC PROTECTOR. IT WAS REPORTED THAT A SELECTIVE CBD (COMMON BILE DUCT) CANNULATION WAS DONE SMOOTHLY, CHOLANGIOGRAM TAKEN THEN AFTER SPHINCTEROTOME WAS POSITION FOR SPHINCTEROTOMY, ON ATTEMPTED SPHINCTEROTOMY CUTTING WIRE DETACHED FROM TIP AND PARTIAL SPHINCTEROTOMY WAS ONLY COMPLETED AND USED ANOTHER SPHINCTEROTOME TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. PER CLARIFICATION RECEIVED ON (B)(6) 2022: CUTTING WIRE ANCHOR SEPARATED FROM CATHETER, BUT DID NOT DETACH. SUBJECT OF REPORT A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
790820 TRI-TOME PC PROTECTOR KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS WILSON-COOK MEDICAL INC W4346930 10827002491974

Patients

Seq Age Sex Outcome Treatment
1 55 YR Male ALLAN CAUTERY ELECTROSURGICAL GENERATOR| BOSTON SCIENTIFIC HYDRA JAGWIRE GUIDEWIRE| OLYMPUS, 160 ENDOSCOPE