THERMAGE FLX (TG-3A) SYSTEM AND ACCESSORIES
Report
- Report Number
- 3011423170-2022-00050
- Event Type
- Malfunction
- Date Received
- May 12, 2022
- Date of Event
- February 3, 2022
- Report Date
- April 12, 2022
- Manufacturer
- SOLTA MEDICAL, INC
- Product Code
- GEI
- PMA / PMN Number
- K132431
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
TREATMENT TIP AND SYSTEM DATA LOGS WERE RETURNED AND EVALUATED. EVALUATION OF THE TREATMENT DATA INDICATES THAT THE SYSTEM AND HAND PIECE PERFORMED AS EXPECTED DURING TREATMENT. THE TIP PASSED THERMISTOR TESTING. THE TIP FAILED THE LEAK TEST. THE TIP FAILED VISUAL INSPECTION DUE TO THE OBSERVANCE OF DIELECTRIC BREAKDOWN BEGINNING TO FORM ON THE RADIO FREQUENCY TRACE. NO DENTS OR SCRATCHES WERE OBSERVED. NO FUNCTIONAL TESTING WAS POSSIBLE DUE TO THE CONDITION OF THE RETURNED DEVICE. A REVIEW OF THE DEVICE MANUFACTURING RECORD IS UNDERWAY.
SERVICE RE-EVALUATED THIS CASE AND IT WAS DETERMINED THAT NO DIELECTRIC BREAKDOWN WAS FOUND ON THE TIP. THE INFORMATION PROVIDED DOES NOT SUGGEST THIS INCIDENT MAY HAVE CAUSED OR CONTRIBUTED TO A DEATH, SERIOUS INJURY OR A SERIOUS DETERIORATION IN STATE OF HEALTH. THIS EVENT IS NO LONGER DEEMED REPORTABLE.
CUSTOMER REPORTED THAT THEIR PATIENT UNDERWENT A THERMAGE FLX TREATMENT AFTER WHICH IT WAS OBSERVED THERE WERE SMALL RED SPOTS ON THE PATIENTS RIGHT SIDE ABDOMEN. IT WAS ALSO NOTED THAT THE SYSTEM DISPLAYED THE ERROR CODE FOR "TIP TOO WARM" DURING THE ABDOMEN TREATMENT. THE CUSTOMER REPORTED THE SPOTS APPEAR TO CORRESPOND TO WHERE THE CENTER OF THE TIP WAS PLACED ON EACH TREATMENT SITE. A REVIEW OF THE CLINICAL DETAILS AND AVAILABLE PATIENT PHOTOS WAS COMPLETED BY A MEDICAL REVIEWER AND IT WAS DETERMINED NO SERIOUS INJURY WAS EXPERIENCED IN THIS CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2132411 | THERMAGE FLX (TG-3A) SYSTEM AND ACCESSORIES | ELECTROSURGICAL,CUTTING & COAGULATION & ACC. | GEI | SOLTA MEDICAL, INC | TT4.00F6-900 | 222 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |