FDA Adverse Event Injury Summary report: N

SYNCHROMED II

MDR report key: 14371862 · Received May 12, 2022

Report

Report Number
3004209178-2022-06102
Event Type
Injury
Date Received
May 12, 2022
Date of Event
May 9, 2022
Report Date
August 24, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LKK
UDI-DI
00763000422615
PMA / PMN Number
P860004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 8780, SERIAL#: (B)(4), IMPLANTED: (B)(6) 2022, EXPLANTED: (B)(6) 2022, PRODUCT TYPE: CATHETER. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 8780, SERIAL/LOT #: (B)(4), UBD: 02-FEB-2024, UDI#: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID 8780 LOT# SERIAL# (B)(6) IMPLANTED: (B)(6) 2022 EXPLANTED: (B)(6) 2022 PRODUCT TYPE CATHETER MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS RECEIVING UNKNOWN BACLOFEN (500MCG/ML AT 55MCG/DAY) VIA AN IMPLANTABLE PUMP FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT PATIENT REPORTED LEAKING FROM BOTH INCISION SITES ABOUT 4 WEEKS AFTER INITIAL IMPLANT AND PROGRESSIVELY RETURNING TO MEDICAL BASELINE OVER A MATTER OF ABOUT A WEEK. NO FACTORS HAVE LED TO THE EVENT. THE PUMP WAS INTERROGATED AND IT SHOWED NO ALERTS. THE CATHETER ACCESS PORT (CAP) WAS WITHDRAWN, WHICH PRODUCE A GOOD STEADY FLOW OF CEREBRA SPINAL FLUID (CSF). THE PUMP POCKET AND CATHETER INSERTION AREA WERE BOTH VISUALIZED FOR ANY EGREGIOUS ABNORMALITIES. THE PUMP SEGMENT OF CATHETER WAS CUT AND REPLACED WITH NEW 8784 SEGMENT. THE DOCTOR DETERMINED TO ALSO REPLACE THE PUMP AT THAT TIME. IT WAS UNKNOWN IF THE ISSUE WAS RESOLVED. THE PATIENT STATUS WAS ALIVE NO INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2311917 SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MEDTRONIC PUERTO RICO OPERATIONS CO. 8637-40 00763000422615

Patients

Seq Age Sex Outcome Treatment
1 29 YR Female Required Intervention "SEE H10...."