TRUE METRIX
Report
- Report Number
- 1000113657-2022-00278
- Event Type
- Malfunction
- Date Received
- May 11, 2022
- Date of Event
- March 22, 2022
- Report Date
- May 11, 2022
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR, US
- Reporter Occupation
- OTHER
Narratives
INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE RETURNED FOR EVALUATION. PRODUCT TEST WAS PERFORMED AND DEFECT FOUND ON RETURNED TEST STRIPS: HIGH RESULTS. NO DEFECT FOUND ON RETURNED METER. ROOT CAUSE: RC-061: STORAGE OUTSIDE SPECIFICATIONS. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON (B)(6) 2022 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 162, 172, 136, 149 AND 143MG/DL. CUSTOMER STATED THE RESULTS USING THE TRUE METRIX METER WERE HIGHER THAN ANOTHER DEVICE (NO METER TO METER COMPARISON RESULTS WERE PROVIDED). THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 102-112MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE LIVING ROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 03/31/2023 AND OPEN VIAL DATE IS ONE MONTH AGO. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1 : 162MG/DL DATE: (B)(6) TIME: 12:44AM FASTING, RESULT 2 : 172MG/DL DATE: (B)(6) TIME: 6:09AM FASTING, RESULT 3 : 136MG/DL DATE: (B)(6) TIME: 5:23AM FASTING, RESULT 4 : 149MG/DL DATE: (B)(6) TIME: 5:23AM FASTING, RESULT 5 : 143MG/DL DATE: (B)(6) TIME: 5:20AM FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1239528 | TRUE METRIX | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TMX AID 50CT24/CASE MG/DL | ZY4495S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |