FDA Adverse Event Malfunction Summary report: N

ARROW CVC SET: 16 GA X 8" (20 CM)

MDR report key: 14349592 · Received May 10, 2022

Report

Report Number
3006425876-2022-00395
Event Type
Malfunction
Date Received
May 10, 2022
Date of Event
January 27, 2022
Report Date
April 18, 2022
Manufacturer
ARROW INTERNATIONAL LLC
Product Code
FOZ
PMA / PMN Number
K900263
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4).

Additional Manufacturer Narrative · 0

(B)(4). COMPLAINT VERIFICATION TESTING COULD NOT BE PERFORMED AS NO SAMPLE WAS RETURNED FOR ANALYSIS. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED WITH ONE RELEVANT FINDING. A NON-CONFORMANCE WAS INITIATED FOR MATERIAL K-04300-019A, LOT 71P21B0956, IN REGARD TO "SYRINGE/NEEDLE CONNECTION NOT SECURE IN USE". THE INSTRUCTIONS FOR USE (IFU) PROVIDED WITH THIS KIT WARNS THE USER, "WARNING: DO NOT ASPIRATE ARROW RAULERSON SYRINGE WHILE GUIDEWIRE IS IN PLACE; AIR MAY ENTER SYRINGE THROUGH REAR VALVE. PRECAUTION: DO NOT REINFUSE BLOOD TO REDUCE RISK OF BLOOD LEAKAGE FROM REAR (CAP) OF SYRINGE." WITHOUT THE DEVICE TO EVALUATE THE COMPLAINT COULD NOT BE CONFIRMED AND THE PROBABLE CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE. IF THE SAMPLE BECOMES AVAILABLE AT A LATER DATE A FOLLOW UP REPORT WILL BE SUBMITTED WITH INVESTIGATION RESULTS.

Description of Event or Problem · 0

IT WAS REPORTED THE ARS (ARROW RAULERSON SYRINGE) WAS FOUND LEAKING DURING USE ON THE PATIENT. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.

Description of Event or Problem · 0

IT WAS REPORTED THE ARS (ARROW RAULERSON SYRINGE) WAS FOUND LEAKING DURING USE ON THE PATIENT. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
871903 ARROW CVC SET: 16 GA X 8" (20 CM) CATHETER,INTRAVASCULAR,THERAP FOZ ARROW INTERNATIONAL LLC 71F21E2412

Patients

Seq Age Sex Outcome Treatment
1 Unknown