SHUNT PAK, 11.5FR DOUBLE-VALVE
Report
- Report Number
- 1625685-2022-00049
- Event Type
- Malfunction
- Date Received
- May 10, 2022
- Date of Event
- April 26, 2022
- Report Date
- July 7, 2022
- Manufacturer
- CAREFUSION, INC
- Product Code
- KPM
- UDI-DI
- 50885403421370
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
(B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).
PR 5064896 FOLLOW-UP EMDR FOR DEVICE EVALUATION: ONE PHOTO AND TWO SEALED PRODUCT TRAYS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, OF THE FIRST TRAY FOREIGN MATTER WAS OBSERVED. THE SECOND TRY WERE EXAMINED TWO 16FR SIZE INTRODUCER WERE OBSERVED; THEREFORE, THE REPORTED FAILURE MODE WAS CONFIRMED. A REVIEW OF THE INTERNAL MANUFACTURING DEVICE RECORDS AND RAW MATERIAL HISTORY FILES FOR THE REPORTED LOT NUMBER 0001453124 WAS PERFORMED AND NO RECORDED QUALITY PROBLEMS OR REJECTIONS RELATED TO THIS INCIDENT WERE FOUND. PRODUCT UNDERGOES INSPECTIONS DURING MANUFACTURING, NO ISSUES RELATED TO THE REPORTED INCIDENT WERE IDENTIFIED, ALL PROCEDURAL AND FUNCTIONAL REQUIREMENTS FOR PRODUCT RELEASE HAVE BEEN MET. BASED ON THE QUALITY TEAM'S INVESTIGATION, PROBABLE ROOT CAUSES WERE DETERMINED AS MANUFACTURING RELATED. MANUFACTURING PERSONNEL HAVE BEEN NOTIFIED OF THIS INCIDENT AND ADDITIONAL TRAINING WAS PROVIDED. COMPLAINTS RECEIVED FOR THIS DEVICE AND DEFECT WILL CONTINUE TO BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS. H3 OTHER TEXT : SEE MANUFACTURE NARRATION.
IT WAS REPORTED BY THE CUSTOMER THAT "THERE IS A LONG HAIR IN THE TRAY" AND " THERE ARE TWO PIECES OF 16FR. INTRODUCERS IN A KIT. AND 12 FR. IS NOT INCLUDED". VERBATIM: THERE IS A LONG HAIR IN THE TRAY. CATALOG NO. 42-2090, LOT 0001453124, QUANTITY 1. WE FOUND OUT DURING INCOMING INSPECTION AT (B)(6). SO THEY ARE NOT USED ON THE PATIENT.
IT WAS REPORTED BY THE CUSTOMER THAT "THERE IS A LONG HAIR IN THE TRAY" AND " THERE ARE TWO PIECES OF 16FR. INTRODUCERS IN A KIT. AND 12 FR. IS NOT INCLUDED". VERBATIM: THERE IS A LONG HAIR IN THE TRAY. (SEE ATTACHED PHOTO) CATALOG NO. 42-2090 LOT 0001453124 QUANTITY 1. WE FOUND OUT DURING INCOMING INSPECTION AT MIHAMA. SO THEY ARE NOT USED ON THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1171404 | SHUNT PAK, 11.5FR DOUBLE-VALVE | NA | KPM | CAREFUSION, INC | 0001453124 | 50885403421370 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |