FDA Adverse Event Malfunction Summary report: N

SHUNT PAK, 11.5FR DOUBLE-VALVE

MDR report key: 14348417 · Received May 10, 2022

Report

Report Number
1625685-2022-00049
Event Type
Malfunction
Date Received
May 10, 2022
Date of Event
April 26, 2022
Report Date
July 7, 2022
Manufacturer
CAREFUSION, INC
Product Code
KPM
UDI-DI
50885403421370
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).

Additional Manufacturer Narrative · 0

PR 5064896 FOLLOW-UP EMDR FOR DEVICE EVALUATION: ONE PHOTO AND TWO SEALED PRODUCT TRAYS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, OF THE FIRST TRAY FOREIGN MATTER WAS OBSERVED. THE SECOND TRY WERE EXAMINED TWO 16FR SIZE INTRODUCER WERE OBSERVED; THEREFORE, THE REPORTED FAILURE MODE WAS CONFIRMED. A REVIEW OF THE INTERNAL MANUFACTURING DEVICE RECORDS AND RAW MATERIAL HISTORY FILES FOR THE REPORTED LOT NUMBER 0001453124 WAS PERFORMED AND NO RECORDED QUALITY PROBLEMS OR REJECTIONS RELATED TO THIS INCIDENT WERE FOUND. PRODUCT UNDERGOES INSPECTIONS DURING MANUFACTURING, NO ISSUES RELATED TO THE REPORTED INCIDENT WERE IDENTIFIED, ALL PROCEDURAL AND FUNCTIONAL REQUIREMENTS FOR PRODUCT RELEASE HAVE BEEN MET. BASED ON THE QUALITY TEAM'S INVESTIGATION, PROBABLE ROOT CAUSES WERE DETERMINED AS MANUFACTURING RELATED. MANUFACTURING PERSONNEL HAVE BEEN NOTIFIED OF THIS INCIDENT AND ADDITIONAL TRAINING WAS PROVIDED. COMPLAINTS RECEIVED FOR THIS DEVICE AND DEFECT WILL CONTINUE TO BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS. H3 OTHER TEXT : SEE MANUFACTURE NARRATION.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT "THERE IS A LONG HAIR IN THE TRAY" AND " THERE ARE TWO PIECES OF 16FR. INTRODUCERS IN A KIT. AND 12 FR. IS NOT INCLUDED". VERBATIM: THERE IS A LONG HAIR IN THE TRAY. CATALOG NO. 42-2090, LOT 0001453124, QUANTITY 1. WE FOUND OUT DURING INCOMING INSPECTION AT (B)(6). SO THEY ARE NOT USED ON THE PATIENT.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT "THERE IS A LONG HAIR IN THE TRAY" AND " THERE ARE TWO PIECES OF 16FR. INTRODUCERS IN A KIT. AND 12 FR. IS NOT INCLUDED". VERBATIM: THERE IS A LONG HAIR IN THE TRAY. (SEE ATTACHED PHOTO) CATALOG NO. 42-2090 LOT 0001453124 QUANTITY 1. WE FOUND OUT DURING INCOMING INSPECTION AT MIHAMA. SO THEY ARE NOT USED ON THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1171404 SHUNT PAK, 11.5FR DOUBLE-VALVE NA KPM CAREFUSION, INC 0001453124 50885403421370

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other