FDA Adverse Event Malfunction Summary report: N

VERIGENE PROCESSOR SP

MDR report key: 14342674 · Received May 9, 2022

Report

Report Number
1650733-2022-00025
Event Type
Malfunction
Date Received
May 9, 2022
Date of Event
January 20, 2022
Report Date
May 9, 2022
Manufacturer
LUMINEX CORPORATION
Product Code
NSU
UDI-DI
00840487101483
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IN REVIEW OF ALL ASSOCIATED CUSTOMER-PROVIDED DOCUMENTATION AND INTERNALLY GENERATED INVESTIGATION NOTES AGAINST THE CRITERIA SET FORTH IN DOCUMENT 00043 (MEDICAL DEVICE REPORTING STANDARDS OPERATING PROCEDURE), THIS COMPLAINT DOES MEET THE REQUIREMENTS FOR AN MDR IN THE COUNTRIES WHERE THIS PRODUCT IS SOLD OR DISTRIBUTED. FILING INDIVIDUAL MDR OUT OF ABUNDANCE OF CAUTION DUE TO THE REMOTE POSSIBILITY OF A FALSE RESULT CONTRIBUTING TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 0

THE SERVICE VISIT AT THE KAISER PERMANENTE REGIONAL LAB WAS SCHEDULED PER CAPA (B)(4)WITHIN WORK ORDER WO-00199891. DURING THE SITE VISIT, THE HYBRIDIZATION (HYB) HEATER REQUIRED RECALIBRATION DUE TO FAILING INITIAL VERIFICATION AT SETPOINTS 1 AND 2. FIELD QUALITY UPDATED RISK ASSESSMENT QUESTION 2 TO 'YES' SINCE THE SETPOINT TEMPERATURES FAILED OUTSIDE THE ACCEPTED LIMITS ESTABLISHED BY THE VERIGENE I ASSAY TEMPERATURE TOLERANCE GUARD BAND RISK REPORT (03838). DATA REVIEW: PER THE VERIGENE ASSAY TEMPERATURE TOLERANCE (GUARD-BAND) STUDY REPORT (03768),THE STUDY EVALUATED ASSAY PERFORMANCE OF VERIGENE® RP FLEX, EP, CDF, BC-GP AND BC-GN ASSAYS AT VARYING CRITICAL TEMPERATURE PARAMETERS AND ESTABLISHED A FUNCTIONAL RANGE OF TEMPERATURE FOR ASSAY TOLERANCE ABOVE AND BELOW THE NOMINAL SET POINTS. IT WAS DETERMINED WITHIN 03838, THAT THE HYB HEATER HAS AN UPPER AND LOWER TOLERANCE OF +/-2 DEGREE CELSIUS FROM THE SETPOINT TEMPERATURE 1 (38.5C) AND SETPOINT 2 (90.0C). IN REVIEWING THE ATTACHED NAPLOGS, SETPOINT 1 FOR THE HYB HEATER HAD AN ADJUSTED TEMPERATURE OF +39.1C, WITHIN THE RANGE ESTABLISHED IN 03838. HOWEVER; SETPOINT 2 HAD AN ADJUSTED TEMPERATURE OF +93.5C, 1.5C HIGHER THAN THE ACCEPTABLE RANGE PER 03838. DUE THE FINDING, RISK ASSESSMENT QUESTION 2 WAS UPDATED TO YES. SITE PERFORMANCE: THE CUSTOMER SITE IS CURRENTLY RUNNING CDF, BC-GP, AND BC-GN VERIGENE ASSAYS. THE HYB HEATER VERIFICATION FAILURE CAN POTENTIALLY IMPACT ASSAY PERFORMANCE RELATED TO THE BC-GN AND BC-GP ASSAY PER 03838. WITHIN AN 18-MONTH TIMEFRAME FROM THE VERIFICATION FAILURE, THE CUSTOMER SITE HAS NOT REPORTED ANY PATIENT SAMPLE FALSE RESULTS ON EITHER ASSAY. CONCLUSION: THE CUSTOMER SITE REPORTED NO FALSE RESULTS FOR THE SP PROCESSOR S/N (B)(4) THERE IS NO INDICATION OF THIS VERIFICATION FAILURE RESULTING IN AN ADVERSE EFFECT TO PATIENT RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1186127 VERIGENE PROCESSOR SP VERIGENE PROCESSOR SP NSU LUMINEX CORPORATION 10-0000-07 16277059 00840487101483

Patients

Seq Age Sex Outcome Treatment
1 Unknown